Hach PHC28101 User Manual Download Page 1

User Manual

DOC022.5

2

.80024

pH Refillable Probe for Low Ionic Strength Samples: Model
PHC28101 or PHC28103

Safety information

Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol on the instrument is referenced in the
manual with a precautionary statement.

Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European
electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to
the user.

Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life
equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.

Specifications

Note: Specifications are subject to change without notice.

Specifications

Details

Probe type

Digital combination probe with refillable double junction reference and built-in
temperature sensor

pH range

pH 0 to 14 

pH resolution

User-selectable stabilization time and resolution—Fast: 0.1, Fast: 0.01,
Medium: 0.01, Slow: 0.01 or Slow: 0.001 

Slope

-59 mV/pH (90 to 110% at 25 °C per Nernstian theoretical value)

Operating temperature range

0 to 50 °C (32 to 122 °F)

Storage temperature range

5 to 40 °C (41 to 104 °F)

Junction

Open

Reference type

Ag/AgCl

Minimum sample depth

15 mm (0.59 in.)

Dimensions

Diameter: 12 mm (0.47 in.)
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)

Cable connection

M12 digital output and connector compatible with HQd meters

Product overview

The PHC281 series probe is a pH refillable probe made for difficult samples with a built-in
temperature sensor (

Figure 1

). The PHC28101 or PHC28103 probe is available with a

1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory use. The pH refillable probe
has a high electrolyte flow rate and a large open junction ideal for difficult samples, such
as Low Ionic Strength (LIS), Ultrapure, dirty samples and samples with high solids
content. The probe is not suitable for use with organic solvents. A 59-mL bottle of 2.44 M
KCl Electrode Filling Solution is included with the probe.

1

Summary of Contents for PHC28101

Page 1: ...combination probe with refillable double junction reference and built in temperature sensor pH range pH 0 to 14 pH resolution User selectable stabilization time and resolution Fast 0 1 Fast 0 01 Medi...

Page 2: ...viscous storage filling solution completely Blot dry with a lint free cloth 4 Remove the protective tape from the filling hole before initial use refer to Figure 1 on page 2 Dispose of the protective...

Page 3: ...ne with a pH above and one with a pH below the expected sample pH For a one point calibration select the buffer nearest to the expected sample pH The calibration is recorded in the probe and the data...

Page 4: ...rect service life time stamp Set the date and time in the meter before the probe is attached If complete traceability is necessary enter a sample ID and operator ID before measurement Refer to the HQd...

Page 5: ...eat steps 2 4 for additional measurements 7 When measurements are done store the probe refer to Storage on page 10 Run a check standard The run check standard feature validates instrument performance...

Page 6: ...question mark until the probe is recalibrated To correct the probe calibration and status indicator calibrate the probe refer to Calibration on page 3 Advanced operation Parameter specific settings c...

Page 7: ...elect Measurement Options and update the settings Option Description Resolution Sets the resolution 0 1pH Fast 0 01pH Fast default 0 01pH Medium 0 01pH Slow or 0 001pH Slow The resolution affects the...

Page 8: ...ration points necessary before a calibration can be completed 1 default 2 or 3 Slope Limit Sets the slope limit 1 to 10 acceptable slope criteria default 5 The slope must fall within set limits for su...

Page 9: ...ed when this option is selected Change check standard options 1 Make sure a probe is connected to the meter 2 Push and select PHC281 Settings 3 Select Modify Current Settings 4 Select Check Standards...

Page 10: ...ute in deionized water 4 Soak the probe in pH 4 buffer for up to 20 minutes then rinse with deionized water 5 Blot dry with a lint free cloth For fats grease and oils 1 Soak the glass bulb in a warm d...

Page 11: ...rer s website Refer to the HQd Series meter manual for specific instructions for the meter model HQd meter does not support IntelliCAL probe Contact a Technical Support Representative Connect a probe...

Page 12: ...ts by changing the PHC281 calibration settings and method 2 Recalibrate the probe 3 Run a check standard to check the performance of the probe Drifting Inaccurate readings Contaminated glass bulb Clea...

Page 13: ...the current method Create a new method with an expanded range Temperature out of range Temperature value is outside of range Make sure that the sample temperature is within the range of the probe Mak...