![Hach PAT700 DS Basic User Manual Download Page 13](http://html1.mh-extra.com/html/hach/pat700-ds/pat700-ds_basic-user-manual_3196958013.webp)
Table 4 Discrete output 1 and 2
Wiring connector 1-2 Terminal Description
Default value
Designation
1
Digital output 1 (+)
TOC alarm (configurable)
DO1+
2
Digital output 2 (+)
Source stream (configurable)
DO2+
3
Common return (–)
Digital output 1 and 2 common
return
DO1/2–
4
12 VDC power Input power supply
+12 V
5
12 VDC power supply – Input power supply
GND
Table 5 Discrete output 3 and 4
Wiring connector 3-4 Terminal Description
Default value
Designation
1
Digital output 3 (+)
Error alarm (configurable)
DO3+
2
Digital output 4 (+)
Valve state (TOC analysis start,
configurable)
DO4+
3
Common return (–)
Digital output 3 and 4 common
return
DO3/4–
4
12 VDC power Input power supply
+12 V
5
12 VDC power supply – Input power supply
GND
Plumbing
Plumb the instrument
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current material safety data
sheets (MSDS) for safety protocols.
and the steps in this procedure to plumb the instrument.
User-supplied items:
• Wrench, 7/16-in.
• Isolation valve
• Tubing, ¼-in. OD polypropylene, PFA or 316 stainless steel for inlet and outlet
1.
Install the isolation valve on the process pipe. Open and close the isolation valve several times to
let water into the system to remove debris.
2.
Push the tube from the isolation valve into the WATER IN port until it stops. Do not crimp or bend
the tubing.
3.
Tighten the compression nut on the WATER IN port by hand. Make a mark on the compression
nut and tube for reference.
Note: For the dual stream model, do step 2 and 3 for the second WATER IN port.
4.
Tighten the compression nut another 1¼ turns.
5.
Push one end of a 10-ft tube into the WATER OUT port until it stops. Do not crimp or bend the
tubing.
6.
Tighten the compression nut on the water outlet side by hand. Make a mark on the compression
nut and tube for reference.
English
13
Summary of Contents for PAT700 DS
Page 2: ...English 3 Français 26 Español 50 Português 74 中文 98 中文 121 日本語 144 한글 167 2 ...
Page 9: ...Figure 2 Installation dimensions English 9 ...
Page 32: ...Figure 2 Dimensions de l installation 32 Français ...
Page 56: ...Figura 2 Dimensiones de instalación 56 Español ...
Page 80: ...Figura 2 Dimensões de instalação 80 Português ...
Page 104: ...图 2 安装尺寸 104 中文 ...
Page 127: ...图 2 安裝尺寸 中文 127 ...
Page 150: ...図 2 取り付け寸法 150 日本語 ...
Page 173: ...그림 2 설치 치수 한글 173 ...
Page 190: ......
Page 191: ......