Action
Interval
Replace antifoam cartridge internal membrane
Yearly
Antifoam pump maintenance
Yearly
Piercing module maintenance
Yearly
Wizards
This option gives an on-screen visual guide through a number of standard maintenance procedures
to perform on the instrument.
Installation
The installation wizard will be invoked automatically the first time the instrument is switched on to help you
complete the installation of the instrument.
Once the instrument has been successfully installed, the
Skip installation wizard on device startup checkbox
should be checked to ensure a normal startup from there onwards.
Maintenance
It is recommended to use the maintenance wizards for maintenance of the instrument. Simply follow the on-
screen instructions available for each of the options.
Antifoam tank filling
When refilling the cartridge, follow the instructions given in
on page 20.
To ensure the performance and reliability of the instrument, it is strongly
recommended to only use the silicone antifoam supplied by Hach Lange (part
number 33156).
EC sensor maintenance
Chemical reactions within the gas sensor requires that the oxygen sensor be
serviced regularly to restore its original sensitivity. A clear sign that a sensor
maintenance is required is when measurements are noticeably less stable than
usual.
For EC sensor maintenance, please also refer to the instruction sheet entitled
A1100 cleaning procedure for 6110 TPA
which is supplied with the instrument.
Goretex filter replacement
It is recommended to replace this filter every week.
Piercing tip replacement
It is recommended to replace these about every six months.
Note: When replacing the tip, use the mounting tool supplied with the installation kit and apply
some Loctite
®
243 sealant (or similar) on the screw thread of the tip before mounting.
Antifoam pump maintenance Refill or replace the antifoam cartridge as required.
Antifoam cartridge preparation
To optimize the performance and reliability of the instrument, it is strongly recommended to only use
the silicone antifoam supplied by Hach Lange (part number 33156). To fill the cartridge, refer to the
steps that follow and
1.
Lower the piercing assembly (refer to
on page 22) for easier access to
the antifoam cartridge connector. Raise the transparent front door.
2.
Push the refill bottle onto the antifoam cartridge connector until the bottle clicks into place.
3.
Squeeze the refill bottle to force as much antifoam as possible out of the bottle and into the
antifoam cartridge.
4.
Maintain pressure on the bottle to avoid any antifoam being sucked back into the bottle. Press on
the lock on the top of the bottle to release it from the antifoam cartridge, and pull the bottle down
to remove it.
20
English