Table 5 Error messages (continued)
Message
Possible cause
Solution
Wiper failed
The wiper is blocked.
The wiper position is not detected.
The wiper is defective.
Examine the measuring path and clean if
necessary.
Do a wiper test.
Contact technical support.
Flash lamp failed
The flash lamp or flash lamp control
is defective.
Contact technical support.
Humidity
There is too much humidity in the
sensor.
Remove the probe from the basin or flow-through
unit. Examine the humidity value in
Diagnostics/Test > Signals.
Contact technical support.
Temperature is out of
range
The temperature in the sensor is too
high.
Check the temperature in Diagnostics/Test >
Signals.
Check the environmental conditions.
Cool the sensor.
Try a different installation location.
Contact technical support.
Table 6 Warning and reminders
Message
Possible cause
Solution
Humidity
The humidity in the sensor is high.
Remove the probe from the basin or flow-
through unit. Examine the humidity value in
Diagnostics/Test > Signals.
Temperature is out of
range
The temperature in the sensor is too
high.
Check the temperature in Diagnostics/Test
> Signals.
Check the environmental conditions.
Cool the sensor.
Try a different installation location.
Contact technical support.
Wiper replacement
The time interval for servicing the wiper
blade has expired.
Replace the wiper blade.
Shaft seals
The time interval for servicing the shaft
seals has expired.
Contact technical support.
Seals
The time interval for servicing the seals
has expired.
Contact technical support.
Service
The time interval for factory service has
expired.
Contact technical support.
Calibration NO3/NO3N
The set calibration interval has expired.
Complete a standard calibration.
Calibration NO2/NO2N
The set calibration interval has expired.
Complete a standard calibration.
Matrix calibration
The set calibration interval has expired.
Complete a matrix calibration.
Section 8 Replacement parts and accessories
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
English
23
Summary of Contents for NT3100sc
Page 2: ...Table of Contents English 3 Fran ais 27 Espa ol 52 Portugu s 78 102 124 147 169 2...
Page 11: ...Figure 3 Bracket installation English 11...
Page 20: ...Figure 8 Replace wiper blade 1 mm and 2 mm Figure 9 Replace wiper blade 5 mm 20 English...
Page 86: ...Figura 3 Instala o do suporte 86 Portugu s...
Page 104: ...2 1 2 1 1 2 1 2 104...
Page 108: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5 mm 5 2 9 1 mm 2 mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 4 9 108...
Page 109: ...3 109...
Page 110: ...4 5 110...
Page 118: ...9 5 mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 118...
Page 126: ...2 1 1 2 1 2 126...
Page 130: ...2 6 2 2 1 NT3100sc NT3200sc 3 5 mm 5 9 2 9 1 mm 2 mm 5 4 LZY261 3 3 1 9 130...
Page 131: ...3 2 3 5 4 3 131...
Page 132: ...4 5 132...
Page 133: ...3 3 SC 6 6 4 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 133...
Page 140: ...8 1 mm 2 mm 9 5 mm 140...
Page 142: ...10 7 5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 UNSTABLE 142...
Page 143: ...5 6 NO3 NO3 N NO3 NO3 N NO3 NO3 N NO2 NO2 N NO2 NO2 N NO2 NO2 N 8 Web 143...
Page 149: ...2 1 1 2 1 2 UV 149...
Page 153: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5mm 5 2 9 1mm 2mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 9 153...
Page 154: ...4 3 154...
Page 155: ...4 5 155...
Page 156: ...3 3 SC 6 6 4 ON 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 156...
Page 163: ...9 5mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 163...
Page 164: ...c 10mL d 8 SC200 SC1000 SC4500 Claros Enter 9 NT3200sc NT3200sc NT3200sc 10 11 12 10 7 5 164...
Page 165: ...5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 6 165...
Page 171: ...5 1 2 PUR HT CE CMIM UKCA FCC ISED 1 2 2 2 1 2 1 1 171...
Page 172: ...2 1 2 UV 2 1 3 2 1 4 ICES 003 Class A Class A 172...
Page 175: ...1 1 3 2 2 6 2 2 1 NT3100sc NT3200sc 3 9 5 5 2 9 1 2 5 4 LZY261 9 175...
Page 176: ...3 3 1 OFF 3 2 3 5 4 176...
Page 177: ...3 177...
Page 178: ...4 5 178...
Page 186: ...9 5 6 4 MSDS SDS UV 1 Sensor setup 2 3 Maintenance 4 Hold 5 6 7 a 10 b 186...