![Hach LPV445.99.02110 User Manual Download Page 23](http://html1.mh-extra.com/html/hach/lpv445-99-02110/lpv445-99-02110_user-manual_3309919023.webp)
Error
Description
Solution
E-12
Low battery
power
Battery power is too low. Replace the
batteries. Refer to
on page 7.
E-13
Parameter load
failure
The memory of the instrument is defective.
Contact technical support.
E-14 followed
by "_.__" or "0"
if no zero was
present
Zero
measurement
invalid
The zero measurement is too low. Use a
sample cell filled with water and try again.
If the error continues, contact technical
support.
E-15 followed
by "_.__"
Absorbance too
high
Identify if there is a light blockage in the
cell holder. Clean the cell holder. Dilute
the sample. Do the test again.
Note: This instrument can not read absorbance
values higher than 3.5 Abs.
E-20
Signal
measurement out
of range
There is too much light on the light
detector. Make sure that the instrument
cap is fully installed on the cell holder. Do
the test again. If the error continues,
contact technical support.
E-21
Signal
measurement
unstable
There is an unstable signal on the light
detector. There is too much or unstable
ambient light. Make sure that the
instrument cap is fully installed on the cell
holder. Do the test again. If the error
continues, contact technical support.
E-22
Hardware error
The electronic system is defective.
Contact technical support.
The following errors can occur immediately after an instrument update.
Error Description
Solution
E-30 No application
There was an error during the application update.
A valid application was not found on the
instrument. Update the instrument again.
E31 Bootloader update
failed
There was an error during the transmission of the
bootloader update. Update the bootloader again.
English
23
Summary of Contents for LPV445.99.02110
Page 50: ...2 1 1 2 1 2 50...
Page 52: ...FCC 1 2 3 2 2 2 3 1 1 LPV445 99 00110 52...
Page 53: ...1 1 DR300 5 2 AAA 6 3 25 mm 10 mL 7 Hach 4 1 cm 10 mL 3 2 53...
Page 54: ...2 1 3 2 2 4 4 1 3 2 Hach 54...
Page 55: ...3 1 LR HR 3 4 2 5 3 5 55...
Page 56: ...4 2 4 4 1 5 2 6 3 7 4 Bluetooth Bluetooth 3 5 24 1 3 3 Hach 56...
Page 57: ...00 00 2 3 6 MSDS SDS 60 57...
Page 58: ...1 LR HR 2 3 4 5 5 6 6 7 8 9 10 5 11 6 12 13 14 58...
Page 59: ...5 1 4 2 25 mm 10 mL 5 3 1 cm 10 mL 6 4 5 6 59...
Page 60: ...6 6 1 1 http www hach com 2 DR300 Search 3 4 Methods Procedures 5 60...
Page 61: ...7 50 50 1 3 2 rCL 01 01 3 4 5 3 8 8 1 SCA SCA SCA SCA 61...
Page 62: ...67 8 1 1 1 2 3 3 SCA 4 a Edit b 5 8 1 2 SCA 1 3 2 SCA 3 OFF 62...
Page 63: ...SCA 8 2 0 2 5 1 00 2 00 4 00 1 00 4 00 60 cal 8 2 1 MSDS SDS 63...
Page 64: ...1 LR HR 2 3 4 a b c 0 00 5 3 6 USEr 7 CAL 8 S0 9 00 00 000 0 10 A0 S0 11 64...
Page 65: ...12 13 14 S1 Add 15 16 A1 a b c S1 17 11 16 18 19 20 3 65...
Page 66: ...8 2 2 10 1 LR HR 2 3 3 USEr 4 Edit 5 S0 6 S0 A0 7 5 6 S1 A1 8 a Add b 5 6 9 3 66...
Page 67: ...8 2 3 1 3 2 USEr 3 dFL 9 9 1 9 2 MSDS SDS 67...
Page 68: ...Liquinox 9 3 53 68...
Page 69: ...10 E 00 E 01 7 57 E 02 LED7 LED LED 7 E 01 E 02 _ __ E 01 E 02 69...
Page 70: ...E 03 E 04 8 E 06 8 10 70...
Page 71: ...E 07 E 09 5 E 10 48 E 12 53 E 13 E 14 _ __ 0 E 15 _ __ 3 5 Abs 71...
Page 72: ...E 20 E 21 E 22 E 30 E31 E 32 E 66 72...
Page 191: ......