Hach LDO10101 User Manual Download Page 9

Section 6 Sample measurement

The procedure that follows is applicable to meters that can connect to Intellical LDO probes. Refer to
the applicable meter documentation for meter operation and probe-specific settings.

6.1 Sample measurement notes

Read the notes that follow before sample measurements.

• Do not submerge the pressure sensor module.
• Salinity changes the solubility of oxygen in water. Measure the sample salinity and enter the value

in the probe settings of the meter.

• To decrease the measurement time, condition the probe and try different stir rates.
• Rinse the probe with deionized water and dry with a lint-free cloth between measurements to

prevent contamination.

• If complete traceability is necessary, enter a sample ID and operator ID before measurement.

Refer to the meter manual for instructions.

• The meter automatically saves the measurement data when the user manually reads each data

point and when the meter is set to read at regular intervals. The user must manually save each
data point when the meter is set to read continuously.

• Air bubbles below the sensor can cause a slow response or error in the measurement. Make sure

to remove air bubbles before and during measurements.

• Make sure to install the shroud before field use to prevent damage to the sensing elements. Refer

to 

Remove or install the shroud

 on page 9. The probe warranty does not include such damage.

• To deploy a rugged probe at a distance, toss the probe body with a slow underhand throw. Do not

throw the probe by the cable to prevent damage to cable, the probe or the user.

6.2 Sample measurement procedure

1. 

 Rinse the probe with

deionized water. Dry
the probe with a lint-
free cloth.
Rugged probes: install
the shroud.

2. 

 Put the probe in the

sample 25 mm (1 inch)
or more. Do not put the
probe on the bottom or
sides of the container.
Stir the sample at a
moderate rate or put
the probe in flowing
conditions.

3. 

 Shake the probe

from side to side to
remove air bubbles.
Rugged probes: move
the probe up and down
to remove air bubbles.

4. 

 Stir gently, then read

the dissolved oxygen
value of the sample.
The display shows the
dissolved oxygen value
when the reading is
stable.

Section 7 Maintenance

7.1 Clean the probe

N O T I C E

 

The LDO sensor cap has a protective black layer to increase the life of the LDO sensor. Do not rub the black layer
to clean the LDO sensor cap. Do not use alcohol or other organic solvents to clean the LDO sensor cap.

Keep the LDO sensor cap clean for best results. Use only water and neutral detergents to clean the
probe.

English

   

7

Summary of Contents for LDO10101

Page 1: ...DOC022 53 80021 LDO101 05 2022 Edition 5 User Manual ...

Page 2: ......

Page 3: ... Preparation for use 5 Section 5 Calibration 6 5 1 Calibration notes 6 5 2 LDO water saturated air 100 calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample measurement notes 7 6 2 Sample measurement procedure 7 Section 7 Maintenance 7 7 1 Clean the probe 7 7 2 Replace the sensor cap and iButton 9 7 3 Remove or install the shroud 9 7 4 Storage 11 Section 8 Troubleshooting 12 Section 9 ...

Page 4: ...Table of Contents 2 ...

Page 5: ...ology and a stirrer for measurements in a BOD bottle Figure 1 Probe overview 1 LDO sensor cap 5 Pressure sensor module 2 Rugged probe 6 iButton 1 2 3 Standard probe 7 iButton compartment 4 Cable 8 Pressure sensor module cap Section 2 Specifications Specifications are subject to change without notice Specifications Details Probe type Luminescent dissolved oxygen LDO probe Dissolved oxygen range 0 1...

Page 6: ...ble connection M12 digital output and connector Warranty 3 years on the probe 1 year on the sensor cap This warranty covers manufacturing defects but not improper use or wear Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water quality parameters in the laboratory or in the field The Intellical probes d...

Page 7: ...are the probe for calibration and measurement as follows Do not touch the protective black layer on the LDO sensor cap 1 Remove the shroud Refer to Remove or install the shroud on page 9 2 Install the LDO sensor cap and tighten by hand Do not tighten more than is necessary for a good seal 3 Soak the probe in 100 mL of tap water for 30 minutes before use For probes that will be in aqueous solutions...

Page 8: ...bration reminders The calibration data is stored in the probe When a calibrated probe is connected to a different meter with the same calibration options a new calibration is not necessary Air bubbles below the sensor when in solution can cause a slow response or error in the calibration Make sure to remove air bubbles during calibration The meter uses the slope value shown at the end of the calib...

Page 9: ...before and during measurements Make sure to install the shroud before field use to prevent damage to the sensing elements Refer to Remove or install the shroud on page 9 The probe warranty does not include such damage To deploy a rugged probe at a distance toss the probe body with a slow underhand throw Do not throw the probe by the cable to prevent damage to cable the probe or the user 6 2 Sample...

Page 10: ...asive cleaners c Install the LDO sensor cap and tighten by hand Do not tighten more than is necessary for a good seal 3 Put the probe in a neutral cleaning solution and stir the solution Do not rub or remove the black layer on the LDO sensor cap Note If disinfection is necessary put the probe in a dilute alcohol solution for a maximum of 2 minutes and rinse immediately with deionized water Do not ...

Page 11: ...5 days or more frequently if the cap becomes damaged or dirty The LDO sensor cap and iButton operate together and must be replaced at the same time Refer to the instructions that are supplied with the LDO sensor cap replacement kit 7 3 Remove or install the shroud Remove the shroud during calibration and maintenance Refer to Figure 3 Keep the shroud installed on the probes during sample measuremen...

Page 12: ...Figure 3 Remove the shroud 10 English ...

Page 13: ...aily measurements in the laboratory keep the probe in a humid environment such as a BOD bottle filled approximately full with tap water If the probe is used to monitor DO concentrations continuously for more than 6 hours keep the probe in tap water When the probe is used for short periods less than 6 hours the storage can be dry Before dry storage rinse the probe in deionized water and blot dry So...

Page 14: ...slow stabilization or inaccurate measurements The measurement was not adjusted for salinity in the sample Measure the salinity of the sample and enter the value as a salinity correction factor in the meter Procedure problem causes slow stabilization and prevents accurate calibrations or measurements Air bubbles are around or below the probe tip Carefully tap or shake the probe to remove air bubble...

Page 15: ......

Page 16: ...rs hach com www hach com HACH LANGE GMBH Willstätterstraße 11 D 40549 Düsseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE Sàrl 6 route de Compois 1222 Vésenaz SWITZERLAND Tel 41 22 594 6400 Fax 41 22 594 6499 Hach Company Hach Lange GmbH 2010 2013 2021 2022 All rights reserved ...

Reviews: