Operation
5.4 Calibration with standard solution
Select
Menu level
Accept
1.
MAIN MENU
2.
SENSOR-SETUP
3.
SELECT SENSOR
(if there is more than one sensor)
4.
CALIBRATION
5.
STANDARD
6.
OUTPUT MODE information
7.
TURN OFF SAMPLE INLET
8.
DRAIN MEAS.CHAMBER
(via the lower feed union)
9.
POUR
STD INTO
MEAS.CHAMBER
(For this purpose connect calibration syringe
to the feed union, see Fig.)
10.
PRESS ENTER
WHEN CONSTANT
x.xxx TRBFNU
11.
CALIBRATE
(x.xxxTRBFNU)
(Enter concentrate of the standard solution)
12.
REMOVE CAL
SYRINGE
13.
CONNECT
SAMPLE INLET
14.
CAL. READY
OPEN INLET
15.
OUTPUT
ACTIVE