Installing Pump/Valve Pinch Plate
1.9 Installing Pump/Valve Pinch Plate
To eliminate the effects of constant pressure on the pump tubes during
shipment and storage, the pinch plate and screws for the pump/valve
module are shipped in the installation kit and the pump tubes are held in place
with tape.
Reagent flow through the pump/valve module must be from bottom to top. If
reversed, fluid mixture will be pumped from the colorimeter sample cell,
causing reagent bottles to overflow.
Complete the assembly of the pump/valve module as follows:
1.
Remove the tape.
2.
Make sure the individual pinch blocks are positioned as shown in
, with the dissimilar-shaped block on top.
3.
Align the pinch plate on the pump/valve module (refer to
4.
Install the two screws through the pinch plate and into the pump/valve
module. Secure the pinch plate by turning the screws in small increments
moving from one screw to the other so that the plate is drawn down
evenly. Tighten until the plate is seated against the pump/valve module.
Figure 15
Installing Pump/Valve Module Pinch Blocks
SAMPLE IN
5 PSIG MAX
1/4IN. O.D. TUBE
AIR PURGE
.01CFM AT
20 PSIG MAX
¼ IN. O.D. TUBE
THIS CLASS A DIGITAL APPARATUS MEETS ALL REQUIREMENTS
OF THE CANADIAN INTERFERENCE CAUSING REGULATIONS.
CET APPAREIL NUMERIQUE DE LA CLASSE “A” RESPECTE
TOUTES LES EXIGENCES DU REGLEMENT SUR LE MATERIEL
BROUILLEUR DU CANADA.
SAMPLE DRAIN
1/2 IN. I.D. TUBE
CASE DRAIN
1/2 IN. I.D. TUBE
ENCLOSURE
115V
J4
F2
F1
GND
NEU
2
Æ
HO
T
T
1
Æ
NO
NO
NC
NC
GND
COM
COM
J9
K1
K2
115V
!
REA
GENT
EXIT
ENTER
Chlorine
mg/L
Alarms
µg/L
MENU
+
B
U
FFER
!
IP 62
GND
F1
F2
50 / 60Hz
AC Voltage
Fuse
F1 & F2
115V / 230 V
T, 2.5A
250V
BLK 1
+
WHT 2
GND
RECORDER
RELAYS
A
C
P
O
W
E
R
CAUTION
ATTENTION
: FOR USE IN A CONTROLLED
ENVIRONMENT. REFER TO MANUAL
FOR ENVIRONMENTAL CONDITIONS.
: EN CAS D’UTILISATION EN
ATMOSPHERE CONTROLEE, CONSULTER
LA NOTICE TECHNIQUE.
COM
NO
NC
1
2
3
COM
NO
NC
1
1
2
2
3
3
#1
#2
Pump/Valve Module
Remove tape from tubes.
Summary of Contents for Lange CL17
Page 1: ...Catalog Number 54400 18 CL17 CHLORINE ANALYZER INSTRUMENT MANUAL 06 04 4ed...
Page 2: ...Hach Company 2001 2004 All rights reserved Printed in the U S A...
Page 10: ...Visit http www hach com...
Page 12: ...Visit http www hach com...
Page 32: ...Visit http www hach com...
Page 46: ...Visit http www hach com...
Page 54: ...Visit http www hach com...
Page 58: ...Visit http www hach com...
Page 62: ...Visit http www hach com...
Page 68: ...Visit http www hach com...