9.
Set the DISPLAY knob to APP E V.
10.
Select a row in
. Turn the APP E knob until the display shows the value in the first column
of the row.
11.
Set the DISPLAY knob to OUTPUT nA.
12.
Set the RANGE nA knob to the value in the second column of the row.
13.
Read across the row to the typical readings. Compare the typical readings to the actual
potentiostat and the multimeter readings.
Note: The multimeter reading can take 10 to 15 seconds to become stable.
14.
If the potentiostat and multimeter readings do not agree with the typical readings in
,
contact BASi technical support.
15.
Set the DISPLAY knob to OFFSET nA.
16.
Turn the OFFSET knob until the display shows "10.0" (10 nA).
17.
If the multimeter reading is not 0 to 150 mV, contact BASi technical support.
Table 3 Typical readings for diagnostic test
App E
Range
Potentiostat reading
Multimeter reading
1.000 V
200 nA
98.0 to 102.0 nA
0.490 to 0.510 V
1.000 V
100 nA
98.0 to 102.0 nA
0.980 to 10.20 V
1.000 V
50 nA
98.0 to 102.0 nA
1.960 to 2.040 V
0.100 V
20 nA
9.50 to 10.50 nA
0.475 to 0.525 V
0.100 V
10 nA
9.50 to 10.50 nA
0.950 to 1.050 V
0.100 V
5 nA
9.50 to 10.50 nA
1.900 to 2.510 V
Replacement parts
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Contact BASi to buy BASi replacement parts. Contact Hach to buy Lachat replacement parts.
Replacement parts—BASi
Description
Quantity
Item no.
Alumina, polishing, 0.05 μm
7 mL
CF-1050
Alumina polish pads
20/pkg
MF-1040
Amperometric detector enclosure
each
CC-5e
Auxiliary electrode
each
MF-1092
Battery, potentiostat, NiCad rechargeable 9V battery
each
ER-9701
Cell gaskets, 51-µM (0.002-in.) thick
4/pkg
MF-1046
Cell lead cable with red clip connector
each
EW-8112
Clamp, quick-release
each
MR-3609 and
MR-3742
Cross-flow cell kit
each
MW-5052
English