Hach IntelliCAL ISENH3181 Series User Manual Download Page 15

Table 1  Troubleshooting information (continued)

Problem

Possible cause

Solution

Procedure problem causes
slow stabilization and
prevents accurate
calibrations or
measurements.

Air bubbles are around or
below the probe tip.

Carefully tap or shake the probe to
remove air bubbles.

The ISA was not added.

Add one ISA powder pillow to each
25 mL of sample and standard
solution.

The stir speed is too slow or
too fast.

Try a different stir speed.

Magnetic stirrers can
become warm and increase
the solution temperature.

Put a piece of insulating material
between the stirrer and beaker.

An incorrect standard
solution was used or the
standard solution has
contamination.

Use the specified standard solution of
good quality.

Too much ammonia gas
goes into the air after the
ISA is added.

Measure the sample or standard
solutions immediately after the ISA is
added to prevent the loss of ammonia
gas. Put a cover on the beaker with a
hole for the probe to decrease the
loss of ammonia.

9.1 Slope check

Use the mV value of two standard solutions to make sure the probe gives the correct slope.

1.

Prepare two standard solutions that are ten times apart in concentration (e.g., 10 mg/L and
100 mg/L NH

3

—N). Select standard solutions with a concentration above and below the typical

sample concentration. Use a minimum concentration of 0.5 mg/L.

2.

Use the measurement procedure to add the ISA and measure the mV value of each standard
solution.

3.

Calculate the difference in the mV value of the two standard solutions to find the slope. If the
probe is in good condition, the slope will be 57 mV (within the ± slope limits of the method) at
25 °C (77 °F).

9.2 Standard additions check

To make sure that the sample measurement is accurate, add a small volume of a standard solution
to the sample and calculate the percent recovery. The sample with the known volume of standard
solution is known as a spiked sample.

1.

Use the measurement procedure to measure the concentration of a 25-mL sample.

2.

Use a pipet to add the applicable volume of standard solution to the sample. Refer to 

Table 2

.

Table 2  Standard solution volumes and concentrations

Measured sample

concentration

Volume of standard to

add

Concentration of standard solution

1 to 2 mg/L

0.5 mL

100 mg/L NH

3

—N

3 to 6 mg/L

1.0 mL

100 mg/L NH

3

—N

7 to 15 mg/L

0.3 mL

1000 mg/L NH

3

—N

English

   

13

Summary of Contents for IntelliCAL ISENH3181 Series

Page 1: ...DOC022 53 80107 ISENH3181 05 2021 Edition 6 User Manual ...

Page 2: ......

Page 3: ... use 5 Section 5 Calibration 6 5 1 Calibration notes 6 5 2 Calibration procedure 7 Section 6 Sample measurement 8 6 1 Sample measurement notes 8 6 2 Sample measurement procedure 8 6 3 Interferences 9 Section 7 Verify the calibration 9 7 1 Verification procedure 10 Section 8 Maintenance 10 8 1 Storage 10 Section 9 Troubleshooting 12 9 1 Slope check 13 9 2 Standard additions check 13 Section 10 Cons...

Page 4: ...Table of Contents 2 ...

Page 5: ...x 5 Electrolyte filling solution 3 Membrane module 3x 6 Glass electrode and temperature sensor Section 2 Specifications Specifications are subject to change without notice Specifications Details Probe type Digital combination gas sensing ammonia ion selective electrode ISE probe with a refillable outer body double junction reference and built in temperature sensor Measurement range 0 01 mg L 5x10 ...

Page 6: ...318103 3 m 9 8 ft Warranty 1 year on the probe This warranty covers manufacturing defects but not improper use or wear Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water quality parameters in the laboratory The Intellical probes do not treat or alter water 3 2 Use of hazard information D A N G E R Ind...

Page 7: ...re hazard Dispose of chemicals and wastes in accordance with local regional and national regulations Section 4 Preparation for use Install a membrane module that contains ammonia electrode filling solution before use 1 Fill and install the membrane module on the probe Do not tighten too much Refer to the illustrated steps in Figure 2 Note Add 12 drops 0 5 mL of the ammonia electrode filling soluti...

Page 8: ...ISA for 30 seconds to 1 minute Keep all of the solutions standard solutions and samples at the same temperature 2 C 3 6 F for best results Stir the standards and samples at a slow and constant rate to prevent the formation of a vortex Use the default calibration options or change the options in the probe settings menu Use the single display mode for calibration when more than one probe is connecte...

Page 9: ...r to the first standard solution Put the standard solution on an electromagnetic stirrer Stir at a moderate rate 5 Rinse the probe with deionized water Dry the probe with a lint free cloth 6 Put the probe in the standard solution with the sensor fully submerged Do not put the probe on the bottom or sides of the beaker 7 Shake the probe from side to side to remove air bubbles 8 Stir for 30 to 60 se...

Page 10: ...all of the solutions standard solutions and samples at the same temperature 2 C 3 6 F for best results If complete traceability is necessary enter a sample ID and operator ID before measurement Refer to the meter manual for instructions Stir the standards and samples at a slow and constant rate to prevent the formation of a vortex The meter automatically saves the measurement data when the user ma...

Page 11: ...ld have about the same concentration of ions and dissolved species as the samples Ammonia forms complexes with a number of metal ions mercury silver copper gold nickel cobalt cadmium and zinc The ammonia ISA contains a high concentration of hydroxide ion and adjusts the pH to 11 where most metals form complexes with hydroxide or precipitate Mercury will form a complex with ammonia when the hydroxi...

Page 12: ...then read the value of the standard solution The meter accepts or rejects the result Section 8 Maintenance Regular maintenance is necessary for the best accuracy stabilization time and life of the probe Make sure to rinse the probe with deionized water between measurements Keep the probe in the recommended storage solution when not in use For dirty applications such as wastewater where the samples...

Page 13: ...not necessary Put a cover on the storage beaker and probe body to prevent solution evaporation Long term more than one week storage Remove the membrane module and soak the probe in pH Electrode Storage Solution Refer to Figure 3 Note After long term storage soak the probe with the installed membrane module in a 1000 mg L NH3 N standard solution without Ionic Strength Adjustor ISA for 30 to 60 minu...

Page 14: ...t the filling solution put the probe with the installed membrane in a 1000 mg L NH3 N standard solution without ISA for 30 seconds to 1 minute The probe is not conditioned sufficiently after membrane replacement Soak the probe with the installed membrane module in a 1000 mg L NH3 N standard solution without ISA for 30 to 60 minutes Sample properties cause slow stabilization or inaccurate measureme...

Page 15: ...ves the correct slope 1 Prepare two standard solutions that are ten times apart in concentration e g 10 mg L and 100 mg L NH3 N Select standard solutions with a concentration above and below the typical sample concentration Use a minimum concentration of 0 5 mg L 2 Use the measurement procedure to add the ISA and measure the mV value of each standard solution 3 Calculate the difference in the mV v...

Page 16: ...here CM measured concentration of the sample after the addition of the standard solution CE expected theoretical concentration of the sample after the addition of the standard solution Section 10 Consumables Note Product and Article numbers may vary for some selling regions Contact the appropriate distributor or refer to the company website for contact information Description Quantity Item no Ammo...

Page 17: ...8042 Probe stand for standard Intellical probes 1 8508850 10 1 Accessories Description Quantity Item no Beaker polypropylene 50 mL low form 1 108041 Disposable wipes 11 x 22 cm 280 pkg 2097000 Wash bottle polyethylene 500 mL 1 62011 Probe stand for standard Intellical probes 1 8508850 English 15 ...

Page 18: ...16 English ...

Page 19: ......

Page 20: ...2932 orders hach com www hach com HACH LANGE GMBH Willstätterstraße 11 D 40549 Düsseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE Sàrl 6 route de Compois 1222 Vésenaz SWITZERLAND Tel 41 22 594 6400 Fax 41 22 594 6499 Hach Company Hach Lange GmbH 2010 2011 2013 2021 All rights reserved ...

Reviews: