Option
Description
Delete standards
Removes a user-specified standard solution from the list of available
standard solutions.
Reminder
Sets a time interval for verification. A reminder shows on the meter after
the time interval. Options: Off (default), 30 minutes, 2 hours, 4 hours,
8 hours, 12 hours or 24 hours.
Acceptance
range
Sets the accepted range of values for the measured verification solution.
Options: 1% to 20% (default: 10%).
Invalidate
calibration
Changes the calibration status to expired if the verification result is not
within the Acceptance range setting. Options: Yes or No (default).
Snooze
reminders
When Reminder is set to a time interval, lets the user ignore a verification
reminder for a selected time. Options: Yes (default) or No.
Snooze time
When Snooze reminders is set to Yes, sets the time when the verification
reminder will show again. Options: 5 minutes, 10 minutes, 15 minutes
(default), 30 minutes or 60 minutes.
6.
Select Probe settings to manage the probe settings as a group.
Option
Description
Current
Selects the probe settings to use for measurements, calibrations and
verifications. Options: Default, Edited or User-specified name. The Edited
option shows when the current settings have changes but are not saved with
a user-specified name.
Create new
settings
Saves the current probe settings with a user-specified name (maximum
16 characters, letters and numbers).
Delete settings
Removes a user-specified probe setting that is not in use.
Section 10 Data management
The meter saves the data from sample measurements, calibrations and verifications as follows:
• Sample measurements—The meter automatically saves the measured sample data when the
measurement mode is Push to read or Interval. When the measurement mode is Continuous, the
user must select Save to save the measured sample data. The save icon shows on the Home
screen when the measured sample data is in the data log.
• Calibration data—The user must select Save to save the calibration data. The calibration data is
saved in the meter and in the Intellical probe.
• Verification data—The user must select Save to save the verification data.
Complete the steps that follow to view, export or delete data.
1.
Push
and select Data, or select Data from the Home screen.
2.
Select an option.
Option
Description
View data
Shows the data in the data log. The most recent data shows first. Push the
down arrow to see more data. The calibration icon shows when the data type
is calibration data. The verification icon shows when the data type is
verification data.
To see more details for a data point, select a data row and push the right
arrow. To apply a filter and show only the filtered data, push the left arrow and
select a parameter, data type or date range.