4.3 Membrane replacement and sensor head cleaning

C A U T I O N

 

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal
protective equipment appropriate to the chemicals that are handled. Refer to the current safety data
sheets (MSDS/SDS) for safety protocols.

A sensor recharge kit is required as it contains all the components necessary for this membrane
replacement and sensor head cleaning process. Refer to 

Sensor recharge kit

 on page 9.

Note: It is advisable to perform this procedure with the plastic sensor base installed so as to avoid any damage to
the connection socket and also to provide a suitable stand for the sensor when required.

1.

Hold the main body of the sensor and unscrew the protection cap locking washer by turning
counter-clockwise. Remove it from the sensor and put to one side.

2.

Pull/twist off the protection cap and put to one side. If you are using a protection cap with a grille,
then remove the Dacron

®

 mesh from inside the cap and discard it.

3.

Hold the sensor with the membrane facing down to avoid spilling any electrolyte, then carefully
unscrew the old cartridge. Drain the old electrolyte into a sink and flush away. Discard the old
cartridge and membrane. Remove the cotton washer and silicone disc from the top of the anode
and discard.

4.

Rinse the sensor head under a tap for 15 seconds to remove any remaining electrolyte and shake
dry. With a soft tissue gently clean around the guard area (indicated right) and then wipe off any
excess moisture from the sensor to ensure all parts are completely dry. Repeat this rinse and dry
process with the protection cap.

5.

Clean the anode using the cleaning tool supplied. Place the tool over the sensor head. Clean by
rotating the cleaning tool over the sensor head for a few seconds, 

in a clockwise direction only

.

6.

Remove the tool and tap it face down on a flat work surface to remove any powdery deposit.
Check the sensor to ensure that all deposits have been removed from the anode. If not, repeat
step 5. until the anode regains its bright silver appearance.

22

   

English

Summary of Contents for GA2 00 Series

Page 1: ...DOC344 52 93153 ORBISPHERE Model GA2X00 O2 EC Sensor 11 2019 Edition 2 User Manual...

Page 2: ......

Page 3: ...Installation 10 3 1 Sensor preparation 10 3 2 Sensor installation 13 3 2 1 Sensor positioning information 13 3 2 2 Sensor insertion 14 3 2 3 Sensor removal 14 3 3 Accessories installation 15 3 3 1 Ext...

Page 4: ...Table of Contents 2...

Page 5: ...erials Stainless steel Stainless steel or hastelloy Sensor head Kalrez O rings Sensor bottom Viton O rings1 EPDM and Kalrez O rings Sensor head Kalrez O rings Sensor bottom Viton O rings or Kalrez O r...

Page 6: ...GA2400 S00 2956A A 33051 S0 Dissolved oxygen traces with 3655 Portable GA2400 S00 2956A A 33051 S0 Table 4 ATEX Sensor Membrane Cartridge Protection Cap GA2800 SVS 2956A A or 29552A A 33051 SG GA2800...

Page 7: ...sec 200 100 30 30 Recommended gaseous flow rate L min 0 1 to 3 1 3 2 Oxygen sensors Table 2 Table 6 Membrane specifications Oxygen sensors 2 Specification 2935A A 29521A A 2995A A Recommended applica...

Page 8: ...product including without limitation direct incidental and consequential damages and disclaims such damages to the full extent permitted under applicable law The user is solely responsible to identify...

Page 9: ...anode immersed in an electrolyte solution which is separated from the gaseous or liquid sample by a gas permeable membrane An electronic circuit is linked to the anode and cathode Through an applied...

Page 10: ...Sensor body 7 Cartridge containing electrolyte and membrane 3 Sensor collar 8 Cotton washer 4 Plastic base 9 Silicone disc 5 Protection cap locking washer 2 3 Product components Make sure that all co...

Page 11: ...l be delivered with each sensor part number 33051 SG or 33051 H0 2 3 3 Sensor recharge kit A recharge kit should have been ordered with the sensor as this will be required to initially make the sensor...

Page 12: ...emicals that are handled Refer to the current safety data sheets MSDS SDS for safety protocols Your sensor has been thoroughly cleaned and tested at the factory before shipment It has been shipped wit...

Page 13: ...it on top of the anode Take a new cotton washer from the recharge kit Hold it between thumb and forefinger and position it on top of the silicone disc Note During this step it is very important to ens...

Page 14: ...nger dry 8 Gently screw the sensor clockwise into the replacement cartridge applying minimum pressure to avoid any damage to the screw threads 9 Continue turning until the cartridge is attached to the...

Page 15: ...urn it clockwise to avoid unscrewing the cartridge 13 Finally screw the protection cap locking washer back into place in a clockwise motion and tighten finger tight 3 2 Sensor installation 3 2 1 Senso...

Page 16: ...rings if product leaks are visible Refer to Figure 4 for sensor installation into a micro volume flow chamber Note Do not twist the sensor when inserting it into a micro volume flow chamber This rota...

Page 17: ...o 5 bar absolute 28117C Pressure sensor 0 to 1 bar absolute Figure 5 External pressure sensor 3 3 2 Weld on stainless steel socket The 29501 weld on sensor socket can be used to install a sensor into...

Page 18: ...nd a pressure of up to 20 bars with the sensor in place or not Refer to Figure 8 Sensor insertion is made by inserting the sensor into the housing and tightening the retaining collar until it stops As...

Page 19: ...housing and tightening the retaining collar until it stops Removal is made by unscrewing the collar and pulling the sensor out Be sure that the sample flow has been turned off before inserting or remo...

Page 20: ...7 in add 100 mm 3 94 in for connection length Table 7 Flow Chamber Orientation Sample Orientation of flow chamber Gaseous or liquid media Vertically with connections down and sensor up Center connecti...

Page 21: ...ss steel plugs provided model 28123 The flow chamber is connected to 6 mm or stainless steel tubing by two Swagelok fittings If necessary copper or plastic tubing with a very low permeability can be s...

Page 22: ...Shown here with gas sensor right pressure sensor center optional second sensor left Section 4 Maintenance W A R N I N G Multiple hazards Only qualified personnel must conduct the tasks described in t...

Page 23: ...r part no 29331 and cloth part no 2934 32301 Electrochemical cleaning and regeneration center see below 40089 Tweezers for maintenance kits DG33303 Cleaning tool for sensor polishing for A110X and C11...

Page 24: ...e then remove the Dacron mesh from inside the cap and discard it 3 Hold the sensor with the membrane facing down to avoid spilling any electrolyte then carefully unscrew the old cartridge Drain the ol...

Page 25: ...filled with water 8 On the clean polishing cloth 2934 do the steps that follow a Add a few drops of water b Spread a little of the polishing powder 29331 to form a grey milky liquid 9 Screw the polis...

Page 26: ...tool over the sensor head Pour enough electrolyte 2959 into the cleaning tool until it completely covers all the electrodes Note For DG33619 regeneration cell screw the regeneration cell on the senso...

Page 27: ...repeat the cleaning process if necessary 18 When cleaning is complete unplug the sensor from the cleaning center and re install the plastic sensor base for the rest of the procedure Remove any remain...

Page 28: ...any visible bubbles in the electrolyte remove them using a stirring motion with the packing component Note The recharge cartridge container will be colored black for all applications except for the 61...

Page 29: ...e center of the protection cap It is very important that the mesh is in the center of the protection cap and covering the entire grille Lower the sensor onto the protection cap making sure not to dist...

Page 30: ...ed sensor operation in relatively high dissolved O2 concentration High O2 concentrations generate deposits more quickly Install a less permeable membrane Turn off the analyzer when sensor is not in a...

Page 31: ...ersonal injury damage to the instrument or equipment malfunction The replacement parts in this section are approved by the manufacturer Note Product and Article numbers may vary for some selling regio...

Page 32: ...Flow chamber in stainless steel 316 for use with for model 28117 and 28117 C pressure sensors 32006 Flow chamber in Delrin for the 3655 portable instrument with one meter of tubing 32007D Flow chamber...

Page 33: ...des o rings cotton washers and cleaning tools 29521A A Recharge kit of 4 pre filled cartridges with premounted 2952A membranes for A1xxx and GA2x00 oxygen sensors Includes o rings cotton washers and c...

Page 34: ...78 x 1 78mm 5 pieces 28093 0 Kalrez O ring for pressure sensor 28117 X 18 78 x 1 78mm 1 piece 28093 2 Nitril O rings for pressure sensor 28117 X 18 78 x 1 78mm 5 pieces 28093 4 EPDM O ring set for sta...

Page 35: ......

Page 36: ...0 669 2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl...

Reviews: