FlowCell FC 48/10 USC
8 English
1.4.2 Installation of the sensor
1.
For correct positioning of the sensor inside the FlowCell, an alignment pin is
supplied that has to be installed at the bottom of the optical path of the sensor.
The pin can easily be tightened and adjusted by using a small coin.
2.
The pin has to be adjusted depending on the path length.
3.
When the pin is adjusted, the sensor can be inserted carefully into the
FlowCell. The pin has to be positioned in the guidance opposite to the
monitoring window and will then be led correctly through the FlowCell. The
sensor has to be inserted until there is a resistance.
90° for 10 mm, 2 mm and 0,3 mm
0° for 5 mm and 1 mm
1
Guidance
1