![Hach FL1500 Basic User Manual Download Page 18](http://html2.mh-extra.com/html/hach/fl1500/fl1500_basic-user-manual_3251897018.webp)
Table 4 Sensor wiring (continued)
Signal
US9000
pHD
2
BL9000
AV9000S
Flo-Dar
3
SVS
Flo-Tote 3
Junction box
for US9000
2 V –
Blue
Black
Green
Green
Green
Black
1 SRQ
—
—
Blue
—
—
—
Installation for Flo-Dar or Flo-Tote sensors
Install the air reference tube and external desiccant to make sure that the pressure transducer in the
sensor operates correctly. The desiccant cartridge prevents damage from moisture and debris.
Moisture and debris can decrease the accuracy of the Flo-Dar and Flo-Tote sensors.
If the instrument is installed in a high-moisture environment, install a desiccant cartridge.
1.
Remove the plug from one of the air reference ports. Refer to
Note: Do not remove the plugs from the air reference ports that are not used.
2.
Push the air reference tube from the sensor into the air reference port.
Note: If the sensor cable is pulled through a conduit, protect the wiring and air tube from damage and
contamination.
3.
Install an external desiccant cartridge on the air inlet port. Refer to
and
on page 30.
Figure 9 Air reference tube and desiccant installation
1
Air inlet
3
Plugs—remove only for air reference tubes
2
Desiccant cartridge
4
Air reference tube from sensor
18
English
Summary of Contents for FL1500
Page 9: ...Figure 3 Mounting dimensions English 9...
Page 37: ...Abbildung 3 Montageabmessungen Deutsch 37...
Page 66: ...Figura 3 Dimensiones de montaje 66 Espa ol...
Page 95: ...Figure 3 Dimensions de montage Fran ais 95...
Page 124: ...Figura 3 Dimensioni di montaggio 124 Italiano...
Page 150: ...3 4 5 2 147 USB 1 1 1 7 x 2 2 8 3 9 x 2 4 USB B 10 x 4 5 USB A USB 11 6 6 mm 150...
Page 151: ...PC PC PC USB B 2 2 1 FL1500 3 AC 2 USB A to B UV 2000 m 2 000 m 20 50 mm 1 3 4 4 151...
Page 152: ...3 152...
Page 153: ...4 1 Unistrut 1 2 Unistrut 1 1 173 153...
Page 154: ...DC AC GFCI GFI PE USB 5 5 1 154...
Page 156: ...3 m 0 82 3 3 mm2 18 12 AWG 300 VAC 70 C 1 154 2 3 AC 4 AC 5 6 1 6 7 AC 6 8 9 10 156...
Page 157: ...6 AC 1 AC AC EU UK AU L N G AC 174 AC AC 157...
Page 158: ...3 DC DC 2 2 AC 8307900 12 VDC 12 VDC DC DC 18 AWG DC 1 154 2 DC 3 4 DC 5 7 3 6 7 8 7 DC 158...
Page 159: ...3 DC DC 12 VDC 12 VDC 1 1 2 154 3 4 8 5 8 4 6 7 8 Flo Dar Flo Tote 161 9 8 10 11 159...
Page 162: ...Web AC AC 300 V 18 AWG 147 1 2 154 3 4 5 7 mm 6 10 5 7 8 9 162...
Page 163: ...10 5 NO COM NC AC DC 11 163...
Page 165: ...6 Diagnostics I O USB 7 7 LED LED 1 IO9000 PC PC PC PC USB B PC PC PC USB PC USB 1 150 169 165...
Page 170: ...147 1 2 3 154 4 5 12 173 6 7 170...
Page 171: ...12 1 DC 3 2 AC 13 171...
Page 172: ...13 172...
Page 178: ...175 USB 1 1 1 7 2x 2 8 3 9 2x 4 USB B 10 4x 5 USB USB A 11 6 6 mm PC PC PC USB B 178...
Page 179: ...2 2 1 FL1500 3 AC 2 USB A B UV 2000 m 6562 ft 2000 m 20 50 mm 0 75 2 0 1 3 4 4 179...
Page 180: ...3 180...
Page 181: ...4 1 1 2 1 1 200 181...
Page 182: ...AC DC GFCI GFI PE USB 5 5 1 182...
Page 185: ...6 AC 1 AC AC EU UK AU L N G AC 201 AC AC 185...
Page 186: ...3 DC 2 2 AC 8307900 12 VDC 12 VDC DC DC DC 18 AWG 1 182 2 DC 3 4 DC 5 7 3 6 7 8 7 DC 186...
Page 187: ...3 DC DC 12 VDC 12 VDC 1 2 182 3 4 8 5 8 4 6 7 8 Flo Dar Flo Tote 189 9 8 10 11 187...
Page 190: ...AC AC 300 V 18 AWG 175 1 2 182 3 4 5 7 mm 0 275 6 10 5 7 8 9 190...
Page 191: ...10 5 NO COM NC AC DC 11 191...
Page 192: ...11 1 3 2 4 Totalizers Totalizers 6 6 I O 192...
Page 193: ...6 I O USB 7 7 LED LED IO9000 PC PC PC PC USB B PC PC PC USB PC USB 1 178 196 193...
Page 198: ...1 2 3 182 4 5 12 200 6 7 12 1 DC 3 2 AC 13 198...
Page 199: ...13 199...
Page 208: ...Rysunek 3 Wymiary monta owe 208 Polski...
Page 237: ...Figura 3 Dimens es de montagem Portugu s 237...
Page 266: ...Slika 3 Mere namestitve 266 Slovenski...
Page 295: ...Figur 3 Monteringsm tt Svenska 295...
Page 324: ...ekil 3 Montaj boyutlar 324 T rk e...
Page 350: ...347 U 1 1 1 Keypad 7 2 2 8 3 9 2 4 B USB 10 4 5 A USB U 11 6 6 mm in PC PC PC B USB 2 350...
Page 351: ...2 1 FL1500 3 2 USB A B UV 2000 m 6562 ft 2000 m 1 20 50 mm 0 75 2 0 in 3 4 4 351...
Page 352: ...3 352...
Page 353: ...4 1 1 2 1 1 371 353...
Page 354: ...GFCI GFI USB 5 5 1 354...
Page 356: ...3 m 10 ft 0 82 3 3 mm2 18 12 AWG 300 VAC 70 C 158 F 1 354 2 3 4 5 6 1 6 7 6 8 9 10 356...
Page 357: ...6 1 L N G 372 357...
Page 358: ...3 2 2 8307900 12 VDC 12 VDC 18 AWG 1 354 2 3 4 5 7 3 6 7 8 7 3 12 VDC 12 VDC 358...
Page 359: ...1 2 354 3 4 8 5 8 4 6 7 8 Flo Dar Flo Tote 361 9 8 10 11 359...
Page 362: ...300 V 18 AWG 347 1 2 354 3 4 5 7 mm 0 275 in 6 10 5 7 8 9 10 362...
Page 364: ...6 I O I O U 7 7 LED LED IO9000 PC PC PC PC B USB PC PC PC U PC U 1 350 367 364...
Page 368: ...8 8 368 X 368 X 369 X Flo Dar Flo Tote 361 X 347 1 2 3 354 4 5 12 371 6 7 368...
Page 369: ...12 1 3 2 13 369...
Page 370: ...13 370...
Page 373: ......