The Pour-Thru cell can be used with Hach chemistries provided a 25 mL sample
is analyzed. Exceptions are noted in
.
The Pour-Thru cell can be used with the following Hach chemistries, but special
care must be taken to rinse the cell with deionized water between samples.
The Pour-Thru cell can also be used to run Nitrogen, Ammonia, Nessler Method
and TKN chemistries. Clean the cell by pouring a few sodium thiosulfate
pentahydrate crystals into the cell. Rinse out the crystals with deionized water.
2.11.1 Cleaning the Pour-Thru Cell
Check the cell occasionally for an accumulation of film on the windows. If the
windows appear dirty or hazy, or if bubbles form in the cell, pour 50 mL of a
detergent solution into the cell. Let it soak for several minutes. Rinse thoroughly
with deionized water. Use a soft cloth to clean the sample cell windows. Paper
towels and other paper products may scratch the windows.
Note:
Do not use solvents (e.g. acetone) to clean the Pour-Thru Cell. A dilute acid solution can be
used for cleaning. Rinse thoroughly with deionized water.
2.12 Using the General-Purpose Timer
Table 1 Hach chemistries that cannot be used with the Pour-Thru cell
Aluminum ECR
Arsenic
Barium
Boron, Carmine
Cyanuric Acid
Fluoride
Formaldehyde
Lead, LeadTrak
Mercury
Nickel, Heptoxime
Nitrite, HR
PCB
Phenols
Potassium
Selenium
Silver
Suspended Solids
Sulfate
TPH
Volatile Acids
Zinc
Surfactants, Anionic (Detergents)
Table 2 Hach chemistries that need extra rinsing with the Pour-Thru cell
Aluminum, Aluminon
Chlorine Dioxide, LR
Cobalt, PAN
Copper, Porphyrin
Hardness, Calmagite
Manganese, LR, PAN
Nickel, PAN
Nitrate, MR
Nitrate, HR
In addition to pre-programmed timers, many programs have a
general purpose timer available. Touch the timer icon to access the
timer system.
Summary of Contents for DR2400
Page 2: ...Page 2 5940018Title fm...
Page 6: ...Page 6 5940018TOC fm...
Page 8: ...Page 8 5940018Safety fm...
Page 10: ...Page 10 5940018Spec fm...
Page 12: ...Page 12 5940018InstStop fm...
Page 20: ...Page 20 5940018Getting Started fm...
Page 32: ...Page 32 5940018Operation fm...
Page 46: ...Page 46 5940018StdAdd fm...
Page 48: ...Page 48 5940018SingleWave fm...
Page 50: ...Page 50 5940018MaintStop fm...
Page 56: ...Page 56 5940018Troubleshoot fm...
Page 58: ...Page 58 5940018GenInfo fm...