General Instrument Operation
General Instrument Operation
Page 28
5940018Operation.fm
Use the optional Pour-Thru Cell (Cat. No. 59404-00) with the Multi-pathlength
Cell Adapter (Cat. No. 59466-00) to save time when analyzing the same
parameter in many samples. Because the same optical characteristics exist for
both zeroing and measuring, or when comparing measurements of different
samples, any error that would have resulted from optical differences between
individual sample cells is eliminated. Rapid Liquid methods speed analysis and
require use of the Pour-Thru Cell (see the
Procedure Manual
for these programs).
When using the Pour-thru Cell, keep the instrument lid open. The light cover on
the cell will ensure accurate readings. See the instructions for the Pour-thru Cell
for more information. Use at least 25 mL of sample to completely flush the cell.
2.9 Assembling the Pour-Thru Cell
The Pour-Thru Cell Assembly (Cat. No. 59404-00) must be assembled and
installed before use.
illustrates the assembled unit.
Figure 10
Pour-Thru Cell Assembly
The kit includes:
•
Pour-Thru Cell
•
Cell Holder
•
Stand Assembly
•
Glass Funnel
•
Stand Pipe
•
1/8” ID plastic tubing (6 ft.)
•
1/4” ID rubber tubing (12 ft.)
•
Instruction Sheet
Stand Pipe
Drain Tube
(1/4" ID Rubber)
Stand Assembly
Glass Funnel
Pour-Thru Cell
Cell Adapter
1/8" ID Plastic Tubing
Stand Pipe and
Funnel Alignment
Summary of Contents for DR2400
Page 2: ...Page 2 5940018Title fm...
Page 6: ...Page 6 5940018TOC fm...
Page 8: ...Page 8 5940018Safety fm...
Page 10: ...Page 10 5940018Spec fm...
Page 12: ...Page 12 5940018InstStop fm...
Page 20: ...Page 20 5940018Getting Started fm...
Page 32: ...Page 32 5940018Operation fm...
Page 46: ...Page 46 5940018StdAdd fm...
Page 48: ...Page 48 5940018SingleWave fm...
Page 50: ...Page 50 5940018MaintStop fm...
Page 56: ...Page 56 5940018Troubleshoot fm...
Page 58: ...Page 58 5940018GenInfo fm...