background image

Option

Description

Program name

Enter a name for the user program, (e.g., the name of the parameter).

Program type

Select single wavelength (for applicable spectrophotometers).

4.

Enter the settings for the user program:

Option

Description

Units

mg/L

Wavelength

520 nm

Concentration resolution

0.001

Chemical form

As

Upper limit

0.220

Lower limit

–0.020

Timer 1

15:00

Timer 2

15:00

Timer 3

15:00

Calibration

Read standards

5.

Select 

CALIBRATION: C = A + BA.

 Push 

EDIT

.

6.

In the Read Standards menu, enter the concentration of the prepared standard
solutions.

7. Prepare the blank solution:

 Use a 25-mL sample cell that only contains unreacted

arsenic absorber solution.

8.

Insert the blank solution into the cell holder and push 

ZERO

.

9.

Insert the first prepared standard solution into the cell holder. Make sure that the
standard solution concentration is selected on the display and push 

READ

.

10.

Continue to measure the remaining standard solutions.

11.

Push 

GRAPH

. Make sure that 

FORCE ZERO

 is off.

12.

If the graph is acceptable select

 DONE>EXIT

.

13.

Save (store) the user program. When the user program is selected in the test
procedure, the calibration curve is used to measure the sample concentration.

Summary of Method

A mixture of zinc, stannous chloride, potassium iodide and hydrochloric acid reduces the
arsenic into arsine gas in a specially equipped distillation apparatus. The arsine goes
through a scrubber that uses a cotton ball soaked with lead acetate for sulfide removal.
Then the arsine goes into an absorber tube that contains a solution of silver
diethyldithiocarbamate in pyridine. The arsenic reacts to form a red complex that is read
colorimetrically. A manual calibration is necessary for this procedure. The measurement
wavelength is 520 nm.

Consumables and replacement items

Required reagents

Description

Quantity/test

Unit

Item no.

Arsenic Standard Solution, 1000-mg/L As

varies

100 mL

1457142

Hydrochloric Acid, ACS

25 mL

500 mL

13449

Lead Acetate Solution, 10%

1 mL

100 mL

1458042

Potassium Iodide Solution, 20%

3 mL

100 mL

1456842

Pyridine, ACS

50 mL

500 mL

1446949

6

Arsenic Silver Diethyldithiocarbamate Method (0.200 mg/L)

Summary of Contents for DR 6000

Page 1: ...Before starting A manual calibration is necessary for this procedure refer to Calibration on page 5 Make a new calibration and update the user program for each new lot of arsenic absorber solution Pre...

Page 2: ...p the preserved samples at room temperature for a maximum of 6 months Before analysis adjust the pH to approximately 7 with 5 N sodium hydroxide solution Correct the test result for the dilution cause...

Page 3: ...ric Acid ACS to the distillation flask 10 Use a serological pipet to add 1 mL of Stannous Chloride Solution to the distillation flask 11 Use a serological pipet to add 3 mL of Potassium Iodide Solutio...

Page 4: ...n the blank sample cell 22 Insert the blank into the cell holder Zero 23 Push ZERO The display shows the non zero intercept that is calculated from the user entered calibration curve 24 Prepare the sa...

Page 5: ...s follows a Use a pipet to add 10 0 mL of the 1000 mg L Arsenic Standard Solution into a 1000 mL volumetric flask b Dilute to the mark with deionized water 2 Prepare three calibration standard solutio...

Page 6: ...FORCE ZERO is off 12 If the graph is acceptable select DONE EXIT 13 Save store the user program When the user program is selected in the test procedure the calibration curve is used to measure the sa...

Page 7: ...s stopper 1 each 1457453 Flask volumetric Class A 200 mL 1 each 1457445 Flask volumetric Class A 500 mL 6 each 1457449 Pipet filler safety bulb 1 each 1465100 Pipet serological 5 mL 1 each 53237 Pipet...

Page 8: ...ANCE PRICE INFORMATION AND ORDERING Call 800 227 4224 Contact the HACH office or distributor serving you www hach com techhelp hach com In the U S A Outside the U S A On the Worldwide Web E mail toll...

Reviews: