Select basic program options
When a program is selected, additional parameter options are available.
1.
Push
Options
to access the option menu.
2.
Select the applicable options.
Option
Description
Start Timer
Select a pre-set timer or set a manual timer to make sure that the steps of an analysis
are correctly timed (e.g., reaction times or wait times can be exactly specified). When the
timer is on, the timer icon is shown on the display. The instrument makes an audible
sound when the time is over.
Pre-set timer
—Push
Start
to start the timer. If a stored
program has more than one timed step, push
Stop>Options>Select>Select
to start the
next timer.
Timer Manual
—Enter the applicable time with the navigation keys and push
Done
. Default = 03:00
Operator ID
The operator ID tag associates readings with an individual operator. Refer to
Sample ID
The sample ID tag is used to associate readings with a particular sample or with a
location. Refer to
on page 13.
%T/Abs/Conc
Switch to % transmittance, absorbance or concentration readings.
Transmittance
reading (%)
—Reads the percent of the original light that passes through the sample and
reaches the detector.
Absorbance readings
—The light absorbed by the sample is read
in absorbance units.
Concentration readings
—The measured absorbance values are
converted into concentration values with the program-specific stored equation.
Advanced
Options
Use the advanced options to specify more parameters. Refer to
Back
Go back to the previous menu.
Select advanced program options
Every program has various additional advanced options from which to select.
1.
Push
Options
to access
Options>Advanced Options
.
2.
Use the advanced options to specify more parameters.
Option
Description
Chemical Form
Select the alternate chemical form and the associated measuring range on some factory-
installed methods.
Reagent Blank
The Reagent Blank correction can be used with some of the factory-installed methods.
Enter the result of a completed test using deionized water as the sample. The blank value
is subtracted from every result to correct for any background color due to reagents. Enter
the blank correction before the Standard Adjust option is used. Complete this correction for
each new lot of test reagents.
Standard
Adjust
Change stored calibration. Complete a test on a known standard at a concentration near
the top of the test range. Use this function to adjust the result to align the standard
concentration.
Dilution Factor
Enter a corrective dilution factor for specific characteristics. The entered number prompt
will be multiplied by the result for the adjustment. For example, if the sample has been
diluted by a factor of 2, set the dilution factor to on and enter 2.
Note: When a dilution is in effect, the dilution icon is shown on the display.
Standard
Additions
Examine the reading accuracy. Refer to the method procedure for more information.
Select the reading mode
Select Single Wavelength, Multi Wavelength or Time Course from the
SETTINGS>Photometer
Functions
menu.
12
English
Summary of Contents for DR 1900
Page 2: ...English 3 Fran ais 16 Espa ol 30 Portugu s 44 58 71 85 98 2...
Page 59: ...10 x 10 mm 1 in 13 mm 16 mm 1 in 1 cm 10 mL III CE 1 EU 2 1 59...
Page 61: ...1 1 4 2 5 3 2 2 1 DR 1900 4 2 4x 5 1 10 mL 3 AA 4 6 61...
Page 62: ...4 AA 4 3 3 1 4 1 1 A 13 10 B 1 C 1 1 cm 10 mL D 16 62...
Page 63: ...4 5 5 6 63...
Page 64: ...6 1 4 2 5 3 6 7 64...
Page 65: ...7 1 8 2 9 OK 3 10 4 11 5 12 6 13 7 TNTplus 1 1 POWER POWER 1 Auto Shutoff 1 2 65...
Page 69: ...MSDS SDS Liquinox 62 3 5 3 5 69...
Page 70: ...70...
Page 74: ...1 1 4 2 5 3 2 2 1 DR 1900 4 2 4 5 1 1 10 mL 3 4 6 74...
Page 75: ...3 4 NiMH 4 3 3 1 1 4 1 1 A 13 mm 10 mm B 1 C 1 1 cm 10 mL D 16 mm 75...
Page 76: ...4 5 5 6 76...
Page 77: ...6 1 4 2 5 3 6 ID ID 7 77...
Page 83: ...MSDS SDS Liquinox 75 Absorbance 3 5 3 5 3 5 Over range Under range Reading Failed 83...
Page 86: ...500 ID GLP ID 10 x 10 mm 1 13 mm 16 mm 1 1 cm 10 mL Class II III CE 1 EU 2 1 86...
Page 88: ...1 1 4 2 5 3 88 2 2 1 DR 1900 4 2 4x 5 10 mL 1 3 AA 4 6 88...
Page 89: ...AA 4 NiMH 4 3 3 1 4 1 1 A 13 mm 10 mm B 1 C 1 1 cm 10 mL D 16 mm 89...
Page 90: ...4 5 5 6 90...
Page 91: ...6 1 4 2 5 3 6 ID ID 7 91...
Page 92: ...7 1 8 2 9 3 10 4 11 5 12 ID 6 AC 13 ID 7 TNTplus 1 1 1 92...
Page 94: ...1 2 03 00 ID ID 96 ID 96 ID ID 95 ID 95 T Abs Conc 94 94 1 2 2 2 94...
Page 96: ...ID ID ID 1 ID 2 ID ID ID ID ID ID ID 50 ID ID MSDS SDS Liquinox 96...
Page 97: ...89 89 3 5 3 5 97...
Page 101: ...DR 1900 340 800 nm AA LCK TNTplus 1 1 1 4 2 5 3 2 2 1 DR 1900 4 2 4x 5 1 1 10 mL 3 AA 4 6 101...
Page 102: ...AA NIMH 3 3 102...
Page 103: ...1 4 1 1 A 13 10 B 1 C 1 1 10 D 16 4 5 103...
Page 104: ...5 6 6 1 4 2 SETTINGS 5 3 6 HOME ID ID 7 104...
Page 110: ...102 3 5 3 5 Optical 110...
Page 111: ......