76
Nickel, PAN (Method 8150)
A = mg/L reading from instrument
B = g sample amount from table
C = mL analysis volume from table
Nitrogen TKN (Method 8075)
*These are guidelines only. See the spectrophotometer procedure manual.
A = mg/L reading from instrument
B = g sample amount from table
C = mL analysis volume from table
Expected Ni
conc. (mg/kg)
Sample
amount (g)
Analysis
Volume (mL)
4-220
0.500
20.0
10-580
0.400
10.0
25-1500
0.300
5.00
200-11000
0.200
1.00
750-45000
0.100
0.50
Expected Nitrogen
conc. (mg/kg)
Sample
amount (g)
Analysis
Volume (mL)
42-2200
0.500*
10.0*
106-5600
0.400*
5.00*
350-18000
0.300*
2.00*
1000-56000
0.200*
1.00*
4200-220000
0.100*
0.50*
A
2500
×
B
C
×
------------------------
mg/kg Total Ni
=
A
75
×
B
C
×
-----------------
mg/kg TKN
=
DIGESTION PROCEDURE FOR SOLIDS,
continued
客服:400 688 9809
网址:www.mamots.com 电话:0431-85178766
Summary of Contents for Digesdahl 23130-20
Page 1: ...400 688 9809 www mamots com 0431 85178766...
Page 11: ...2 400 688 9809 www mamots com 0431 85178766...
Page 17: ...8 Figure 3 Fractionating Column and Exhaust Assembly 400 688 9809 www mamots com 0431 85178766...
Page 19: ...10 400 688 9809 www mamots com 0431 85178766...
Page 36: ...27 DIGESTION PROCEDURES 400 688 9809 www mamots com 0431 85178766...
Page 43: ...34 400 688 9809 www mamots com 0431 85178766...
Page 45: ...36 SAMPLE TYPE AND SIZE continued Chart 2 For Oils 400 688 9809 www mamots com 0431 85178766...
Page 60: ...51 400 688 9809 www mamots com 0431 85178766...
Page 74: ...65 400 688 9809 www mamots com 0431 85178766...
Page 89: ...80 400 688 9809 www mamots com 0431 85178766...
Page 91: ...82 400 688 9809 www mamots com 0431 85178766...
Page 93: ...84 400 688 9809 www mamots com 0431 85178766...
Page 96: ...87 HOW TO ORDER 400 688 9809 www mamots com 0431 85178766...
Page 97: ...88 REPAIR SERVICE 400 688 9809 www mamots com 0431 85178766...
Page 98: ...89 WARRANTY 400 688 9809 www mamots com 0431 85178766...
Page 99: ...400 688 9809 www mamots com 0431 85178766...