Hach CDC401 Series User Manual Download Page 1

DOC022.53.80022

CDC401

05/2021, Edition 4

User Manual

Summary of Contents for CDC401 Series

Page 1: ...DOC022 53 80022 CDC401 05 2021 Edition 4 User Manual...

Page 2: ......

Page 3: ...4 Preparation for use 5 Section 5 Calibration 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 6 6 1 Sample measurement notes 6 6 2 Sample measurement procedure 7 Sec...

Page 4: ...Table of Contents 2...

Page 5: ...uctivity cell 3 Shroud rugged model 2 Standard probe 4 Rugged probe Section 2 Specifications Specifications are subject to change without notice Specifications Details Probe type Graphite 4 pole condu...

Page 6: ...0 m 98 4 ft Warranty 1 year on the probe This warranty covers manufacturing defects but not improper use or wear Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical...

Page 7: ...the meter Note Some meters automatically open the date and time settings when the meter starts for the first time or after battery replacement 3 Connect the probe to the meter Section 5 Calibration T...

Page 8: ...put the probe on the bottom or sides of the beaker 5 Shake the probe from side to side to remove air bubbles 6 Stir gently then read the conductivity value of the standard solution The display shows t...

Page 9: ...warranty does not include such damage To deploy a rugged probe at a distance toss the probe body with a slow underhand throw Do not throw the probe by the cable to prevent damage to cable the probe o...

Page 10: ...ndard solution The meter accepts or rejects the result Section 8 Maintenance 8 1 Clean the probe Clean the probe regularly to remove mineral or sample buildup on the electrodes Symptoms of contaminati...

Page 11: ...oils and grease contamination KATHON CG TRITON X 2 hours maximum Proteins Electrode cleaning solution for proteins organic contamination Pepsin in HCl 3 hours maximum Wastewater and organic compounds...

Page 12: ...asurements The sample absorbs carbon dioxide CO2 from the air which forms carbonic acid and slowly increases the conductivity in low ionic strength LIS or high purity samples Use the LIS chamber for L...

Page 13: ...0 pkg 2771320 Conductivity standard solution KCl 1413 S cm Singlet one use packets 20 mL each 20 pkg 2771420 Conductivity standard solution KCl 12 88 mS cm Singlet one use packets 20 mL each 20 pkg 27...

Page 14: ...12 English...

Page 15: ......

Page 16: ...2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 ro...

Reviews: