![Hach AutoCAT 9000 Instruction Manual Download Page 21](http://html1.mh-extra.com/html/hach/autocat-9000/autocat-9000_instruction-manual_3154724021.webp)
Operating Environment
50081_Assembly.fm
Page 19
2.2.6 Installing the Electrode
Figure 3
AutoCAT 9000 Workstation Cable Connections
2.2.7 Power Connection
A UL/CSA-approved 125-V ac power cord (Cat. No. 18010-00) is supplied with the North
American AutoCAT 9000 models.
To power North American AutoCAT 9000 models with 230 V ac, replace the supplied 125-V
ac power cord with a UL/CSA approved 230-V ac power cord. See section
Selecting the ac Power Frequency
for configuration settings.
Plug the power cord into the back panel of the instrument and connect the power cord to the
proper outlet. See
.
After assembly is complete, perform the steps in Section
to enter basic user
information in preparation for performing analyses to comply with Good Laboratory
Practices. Refer to Section
for the software setup required prior to
performing analyses. The software setup in Section
must be completed before any
analyses can be performed.
1.
Remove the protective cap from the electrode.
2.
Insert the electrode into a large hole (opposite the plastic anti-diffusion
tip) on the electrode head.
3.
Turn the electrode ¼-turn clockwise to lock the collar in place.
4.
Connect the cable to the BNC connector on the back of the instrument
(see
Note:
The electrode
must
be installed opposite the plastic anti-diffusion tip to
allow maximum mixing of the titrant before the reaction is sensed
by the electrode.
Legend
Number
Item Name
1
PC/Printer
2
Local
3
Electrode
4
Temperature Sensor
5
Power
6
Fuse Access
7
ON/OFF Switch
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......