Hach AS950 - Quick Start Guide 

 
Time Based Sampling Program  
 
Initial Setup of New Sampler 
 

1.

 

Apply power to the sampler and turn on the controller. 

2.

 

Press MENU key on keypad 

3.

 

Navigate to General Settings 

4.

 

Select Date and Time 

5.

 

Enter the current date and time and select OK 

6.

 

Press MENU key on keypad 

7.

 

Navigate to Hardware Setup 

8.

 

Select Sampler 

9.

 

Select Rinses 

enter “1”

 and select OK 

10.

 

Select Sample Retries enter “1”

 and select OK 

11.

 

Press MENU key on keypad 

12.

 

Navigate to Programming 

13.

 

Select Programming 

14.

 

Navigate to and select Sampler Programming 

15.

 

Select Total Bottles enter “1” and select OK

 

16.

 

Select Bottle Volume enter “3 gallon” and select OK

 

17.

 

Select Tubing enter “

length of intake tubing to the nearest foot

” and select OK

 

18.

 

Select Pacing then select Time Weighted, select NEXT 

19.

 

With Time Weighted highlighted select Edit, enter “

10 minutes

” select OK

 

20.

 

Select Take First. Select Edit, select 

Immediately

 select Back 

21.

 

Select Sample Volume, select Fixed 

22.

 

Select Volume enter “

50 

ml”, select

 OK, select Back 

23.

 

Select Program Start, select Immediately on ‘Run’, select Next

 

24.

 

Select Program End, select Edit, select After Samples, press Select 

25.

 

Highlight Samples, select Edit, enter “1

44

” select OK, select Back

 

26.

 

Select RUN/HALT key on keypad 

27.

 

Select Start Program 

28.

 

If warned about clearing data, affirm by selecting OK 

29.

 

Sampler is now Running 
 

NOTE: 

Bold font

 in step 17 indicates a variable parameter that is site dependent. Each time the 

sampler is moved to a new location you will want to enter the length of tubing between the intake 
strainer and the sampler to the nearest foot. Entering the tubing length ensures that the sample 
volume will be within tolerance. 
 