PAT700 Total Organic Carbon Analyzer – Bottle Mode
Page 77
Anatel
Operator Manual
Grab sample button
takes you to the grab sample load bottles dialog box. See
Excursion mode button
takes you to the excursion mode dialog box. See
Unload bottles button
takes you to the unload/replace bottles dialog box. See
6.4
Run standards
The run standards dialog box allows you to select a standards test to run. See
.
Figure 6-2 Run standards dialog box
TOC calibration button
takes you to the TOC calibration setup dialog box.
Conductivity calibration button
takes you to the conductivity calibration load bottles dialog
Conductivity calibration, page 85
.
TOC validation button
takes you to the TOC validation setup dialog box. See
System suitability button
takes you to the system suitability load bottles dialog box. See
System suitability test, page 91
.
Return icon
returns you to the bottle mode dialog box. See
Bottle mode dialog box, page 76
.
6.5
TOC calibration
TOC calibration ensures the accuracy of the TOC values obtained by the analyzer. TOC
values are determined by measuring the change in conductivity of a high-purity water sample
as its organic compounds oxidize to carbon dioxide inside the analysis cell. The procedure
subtracts the average of a series of background TOC measurements from the average of a
series of user-selected standard measurements (in the form of sucrose). Linear regression is
performed on the data, and the correlation coefficient (R
2
) and percent slope change (% sc)
are calculated and reported.