WG
M - 2
8
No
vemb
e
r 20
07
- Ed
it
io
n 0
1
PAT700 Total Organic Carbon Analyzer – Run Modes
Page 41
Anatel
Operator Manual
Chapter 4
Run Modes
4.1
The analysis cycle
An analysis cycle is comprised of a flush of the analysis cell, oxidation, and idle time (if the
cycle time is greater than the combined flush and oxidation time).
Oxidation time varies based on the amount of total organic carbon in the sample. Flush time
is configurable and idle time varies to allow for varying analysis cycle time. See
.
Idle time occurs only after an online TOC analysis has been completed and only if the elapsed
time has not exceeded the value for cycle time.
During idle time, the analyzer can operate in one of three ways: flow without conductivity, flow
with conductivity, or no flow.
•
When flow without conductivity is selected, the sample valve opens during idle,
allowing continuous flow through the cell. No conductivity is reported.
•
When flow with conductivity is selected, the sample valve opens during idle, allowing
continuous flow through the cell. Conductivity of the water is reported.
•
When no-flow is selected, the analyzer sample valve closes, stopping all flow through
the sample cell during idle.
Figure 4-1 The TOC analysis cycle
Cycle time
(configurable)
Start
Flush time
(configurable)
Oxidation time
(variable)
Result reported
Idle time
(variable)
Summary of Contents for ANATEL PAT700
Page 1: ...ANATEL PAT700 ON LINE TOTAL ORGANIC CARBON ANALYZER Operator Manual...
Page 2: ......
Page 4: ......
Page 156: ......
Page 157: ......