Installation
4.3.2 Installing the Reference Electrode
1.
Place the seven small white Teflon
®*
beads through the hole at the top of the
measuring cell as shown in
. The beads provide better mixing and
self-cleaning.
2.
Remove the reference electrode from its packaging.
3.
Remove the red cap from the electrode.
4.
Connect the male end of the electrode to the female end of the red connector
on the measuring cell.
5.
Pull the electrode from the electrode storage bottle.
6.
Screw the electrode into the hole at the top of the measuring cell.
Note:
Replace the reference electrode yearly. See
.
Figure 6
Installing the Teflon Beads
*Teflon is a registered trademark of Dupont Corporation.
1.
Reference Electrode (Cat. No. 368429,00000)
3.
Teflon Beads (Cat. No. 588801,75008)
2.
Reference Electrode Connector (Cat. No. 09186=A=0500)
4.
Measuring Cell (Cat. No. 09186=A=0100)