![Hach 9186 Instrument Manual Download Page 43](http://html1.mh-extra.com/html/hach/9186/9186_instrument-manual_3179647043.webp)
Modbus Communication
43
8.5 Register Information
The fields shown in the register map are as follows:
Register
—Modbus register number. Note that each register is 16 bits in length.
Floating point registers occupy two back-to-back 16-bit registers.
Data Type
—Representation of the data. Can be floating point, integer,
bit or string.
Length
—The number of 16-bit registers.
R/W
—R for read only or R/W for read/write.
Units
—A number corresponding to the type of units. The names of each unit are
shown in
. An “Rxxxxx” indicates the units are selectable,
and are located in register number xxxxx.
Description
—A description of the register.
8.5.1 Register Map
Table 21 Register Map
Register
Data Type
Length
R/W
Units (U)
Description
40000
Float
2
R
R40075
Concentration measurement
40002
Float
2
R
R40076
Temperature measurement
40004
Float
2
R
27
Intensity measurement
40006
Float
2
R
28
Flow rate measurement
40008
Float
2
R
R40075
Concentration of calibration offset
40010
Float
2
R
R40075
Concentration of calibration gain
40012
Float
2
R
R40076
Temperature calibration offset
40014
Float
2
R
—
Reserved for future application
40016
Float
2
R
25
pH calibration gain
40018.1
Bit
1
R
—
Instrument alarm (programmable)
40018.2
Bit
1
R
—
Instrument alarm (programmable)
40018.3
Bit
1
R
—
Instrument alarm (programmable)
40018.4
Bit
1
R
—
Instrument alarm (programmable)
40019
Integer
1
R
—
MONEC status
40020
Integer
1
R/W
—
Alarm 1 type (enum)
40021
Float
2
R/W
—
Alarm 1 setpoint
40023
Integer
1
R/W
—
Alarm low/high
40024
Integer
1
R/W
15
Alarm 1 delay
40025
Integer
1
R/W
10
Alarm 1 hyst
40026
Integer
1
R/W
—
Relay 1 characteristic
40027
Integer
1
R/W
—
Alarm 2 type (enum)
40028
Float
2
R/W
—
Alarm 2 setpoint
40030
Integer
1
R/W
—
Alarm 2 low/high
40031
Integer
1
R/W
15
Alarm 3 delay
40032
Integer
1
R/W
10
Alarm 3 hyst