1.
From the USER TABLE option select P2>INSERT POINT.
2.
Enter the conductivity (X) as
250.0
and push
enter
.
3.
Enter the concentration% (Y) as
15.0
and push
enter
.
4.
Select SAVE CHANGES? and push
enter
.
The table will now be defined as:
Conductivity
Concentration%
P1
0
0
P2
250
15
P3
750
30
Additional points can be inserted up to a maximum of P10.
Note: The conductivity value taken into account is temperature compensated according to the temperature
compensation option configured.
Temperature tables
The process is the same as described for the concentration tables, but the X value refers to
temperature and the Y value is a factor used to multiply against the non-compensated conductivity
value to obtain a temperature compensated conductivity value.
For example, if the temperature table was configured as follows:
Temperature
Value
P1
0
1
P2
20
1.1
P3
40
1.2
This can be displayed graphically as:
This shows that a temperature of 10° corresponds to a factor of 1.05. Therefore, a non-compensated
conductivity value of 10 mS/cm at 10° will give a temperature compensated value of 10.5 mS/cm (i.e.
10 mS/cm multiplied by a factor of 1.05).
Calibrate the sensor
About sensor calibration
The sensor characteristics slowly shift over time and cause the sensor to lose accuracy. The sensor
must be calibrated regularly to maintain accuracy. The calibration frequency varies with the
application and is best determined by experience.
12
English
Summary of Contents for 8332
Page 2: ...English 3 Français 22 Español 42 中文 64 2 ...
Page 7: ...English 7 ...
Page 8: ...8 English ...
Page 26: ...26 Français ...
Page 27: ...Français 27 ...
Page 46: ...46 Español ...
Page 47: ...Español 47 ...
Page 68: ...68 中文 ...
Page 69: ...中文 69 ...
Page 81: ......