background image

Changes or modifications to this equipment not expressly approved by the party responsible for
compliance could void the user's authority to operate the equipment. This equipment has been tested
and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference, in which case the user will be required to
correct the interference at their expense. The following techniques can be used to reduce
interference problems:

1.

Disconnect the equipment from its power source to verify that it is or is not the source of the
interference.

2.

If the equipment is connected to the same outlet as the device experiencing interference, connect
the equipment to a different outlet.

3.

Move the equipment away from the device receiving the interference.

4.

Reposition the receiving antenna for the device receiving the interference.

5.

Try combinations of the above.

Product overview

D A N G E R

 

Electric shock hazard. The power supply must be connected to an electrical outlet that provides an
earth ground connection.

W A R N I N G

 

Explosion hazard. The 6631800 and 6632500 power supplies are not approved for use in hazardous
locations as defined in the National Electrical Code.

This instruction sheet applies to the following power supplies: 6631800 and 6632500.

Prerequisites:

• Make sure that the AC power source has sufficient capacity for the power supply.
• Make sure that all electrical installations and connections are in accordance with national and local

electrical codes.

• Make sure that the AC power source is grounded and protected with a Ground Fault Current

Interrupt (GFCI).

• If the GFCI breaker in the AC power source fails (opens), determine the cause before restoring

power.

• Do not attempt to make any connection or otherwise handle the electrical components of the unit

when connected to the AC power source if the immediate area is wet, or if hands or clothing are
wet.

• Before starting any maintenance task, disconnect the unit from the AC power source.

Installation

D A N G E R

 

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.

Mounting

Attach the unit to a wall with four #6 self-tapping screws. Refer to 

Figure 1

.

English

   

5

Summary of Contents for 6632500

Page 1: ...DOC276 97 80120 SELV power supply 04 2012 Edition 2 User Instructions Instructions d utilisation Instrucciones para el usuario...

Page 2: ...English 3 Fran ais 8 Espa ol 13 2...

Page 3: ...he right to make changes in this manual and the products it describes at any time without notice or obligation Revised editions are found on the manufacturer s website Safety information N O T I C E T...

Page 4: ...en noted on the product identifies the location of a fuse or current limiting device Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 Au...

Page 5: ...the receiving antenna for the device receiving the interference 5 Try combinations of the above Product overview D A N G E R Electric shock hazard The power supply must be connected to an electrical...

Page 6: ...2 Insert the power cable from the SELV power supply through the strain relief fitting in the bottom of the Flo Station enclosure 3 Install the negative wire black into the left side of the power termi...

Page 7: ...wer 1 DC cable from SELV power supply 3 Negative wire 2 DC power terminal 4 Positive wire Maintenance Cleaning the case 1 Disconnect AC power to the power supply 2 Clean the outside of the case with a...

Page 8: ...n d faut ou d une omission dans ce manuel Le constructeur se r serve le droit d apporter des modifications ce manuel et aux produits d crits tout moment sans avertissement ni obligation Les ditions r...

Page 9: ...une prise connectez la terre positive cette borne le kit de cordon am ricain contient une connexion la terre Ce symbole indique la pr sence de dispositifs sensibles aux d charges lectrostatiques ESD e...

Page 10: ...non la source des perturbations 2 Si l quipement est branch sur le m me circuit de prises que l appareil qui subit des interf rences branchez l quipement sur un circuit diff rent 3 loigner l quipement...

Page 11: ...e pour c bles neutres et de ligne Fixez tous les raccords des c bles l aide d un tournevis 1 Retirez le capot inf rieur du Flo Station 2 Ins rez le c ble d alimentation de l alimentation lectrique SEL...

Page 12: ...de l alimentation lectrique SELV 3 Fil n gatif 2 Borne d alimentation CC 4 Fil positif Maintenance Nettoyage du bo tier 1 D branchez l alimentation CA de l alimentation lectrique 2 Nettoyez la face e...

Page 13: ...fecto u omisi n en este manual El fabricante se reserva el derecho a modificar este manual y los productos que describen en cualquier momento sin aviso ni obligaci n Las ediciones revisadas se encuent...

Page 14: ...producto identifica la ubicaci n de un fusible o de un limitador de corriente El equipo el ctrico marcado con este s mbolo no se podr desechar por medio de los sistemas europeos p blicos de eliminaci...

Page 15: ...ndo la interferencia 4 Cambie la posici n de la antena del dispositivo que recibe la interferencia 5 Trate combinaciones de las opciones descritas Descripci n general del producto P E L I G R O Riesgo...

Page 16: ...es de los cables 1 Quite la tapa inferior de la Flo Station 2 Inserte el cable de corriente de la fuente de alimentaci n SELV a trav s del pasacables de la parte inferior de la carcasa de Flo Station...

Page 17: ...fuente de alimentaci n SELV 3 Cable negativo 2 Terminal de alimentaci n CC 4 Cable positivo Mantenimiento Limpieza de la carcasa 1 Desconecte la alimentaci n CA de la fuente de alimentaci n 2 Limpie l...

Page 18: ......

Page 19: ......

Page 20: ...ma Flow Products in France and the UK FLOWTRONIC SA Rue J H Cool 19a B 4840 Welkenraedt Belgium Tel 32 0 87 899 799 Fax 32 0 87 899 790 www flow tronic com France and UK Sigma Flow Products Only HACH...

Reviews: