Maintenance
6.4 Replacing the Sensor Assembly
The sensor assembly consists of a pre-installed membrane, electrolyte, and electrode.
1.
Unscrew the sensor guard or optional lock ring and remove the spent or damaged
sensor assembly.
2.
Install the o-ring onto the groove in the probe body, see item 2 and 3 in
.
3.
Align the sensor assembly connector with the probe body socket, see item 6 in
.
4.
Reinstall the sensor guard or lock ring.
5.
Condition the new sensor before taking measurements. See
.
Figure 13
Replacing the Sensor Assembly
1.
Probe Body
4.
Sensor Assembly
2.
Item 3 seats in groove.
5.
Sensor Guard (if sensor guard is not used, install the optional
lock ring)
3.
O-ring, (Cat. No. 1857100)
6.
Check for connector orientation during assembly.
2
6
3
1
4
5