Option
Description
OVERFEED TIMER
Sets the maximum amount of time to get to the process setpoint. When the time
expires and the relay does not show the setpoint, the relay is set to off. After an
overfeed alarm occurs, reset the timer manually.
OFF DELAY
Sets a delay time to set the relay to off (default = 5 seconds).
ON DELAY
Sets a delay time to set the relay to on (default = 5 seconds).
5.
Select EVENT CONTROL in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
Description
SET SETPOINT
Sets the value to set the relay to on.
DEADBAND
Sets a delay so that the relay will be stable as the process value converges to the setpoint.
OnMax TIMER
Sets the maximum time the relay stays on (default = 0 min).
OffMax TIMER
Sets the maximum time the relay stays off (default = 0 min).
OnMin TIMER
Sets the time the relay stays on, independent from the measured value (default = 0 min).
OffMin TIMER
Sets the time the relay stays off, independent from the measured value (default = 0 min).
6.
Select SCHEDULER in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
Description
HOLD OUTPUTS
Holds or transfers outputs for the selected channels.
RUN DAYS
Selects the days the relay stays on. Options: Sunday, Monday, Tuesday, Wednesday,
Thursday, Friday, Saturday
START TIME
Sets the start time.
INTERVAL
Sets the time between activation cycles (default = 5 min).
DURATION
Sets the period of time the relay is set to on (default = 30 sec).
OFF DELAY
Sets the time for additional hold/output time after the relay is set to off.
7.
Select WARNING in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
Description
WARNING LEVEL
Sets the level for the warning activation and starts applicable individual warnings.
8.
Select PROCESS EVENT in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
Description
MEASURING 1
Sets the relay to on when cell 1 is measuring.
MEASURING 2
Sets the relay to on when cell 2 is measuring.
CLEANING
Sets the relay to on when a cleaning cycle is started.
ZERO CAL
Sets the relay to on when a zero calibration has started.
SLOPE CAL
Sets the relay to on when a slope calibration has started.
SHUTDOWN
Sets the relay to on when a shutdown cycle is started.
STARTUP
Sets the relay to on when a startup cycle is started.
GRAB SAMPLE
Sets the relay to on when a grab sample cycle is started.