![Hach 5500sc PO4 3-HR Operations Download Page 15](http://html1.mh-extra.com/html/hach/5500sc-po4-3-hr/5500sc-po4-3-hr_operations_3147064015.webp)
View data
The analyzer stores a maximum of 18,000 data points. After 18,000 data points are stored, the oldest
data points are overwritten with new data.
1.
Select VIEW DATA.
2.
Select an option.
Option
Description
ANALYZER DATA
Shows analyzer status information (refer to
).
MEASUREMENT DATA
Shows measurement information (refer to
LOG DATA
Selects the data log and/or the event log. DATA LOG—shows the measurement
values. Select start time, number of hours and/or number of readings. EVENT LOG
—shows all analyzer information (e.g. alarms, warnings, configuration changes,
and so on.). Select start time, number of hours and/or number of readings.
Table 2 Analyzer data
Element
Definition
CELL TEMP
Temperature of the colorimeter cell block heater (ideally 49.8 °C to 50.2 °C (121.64 °F
to 122.36 °F))
REAGENT TEMP
Reagent temperature before it enters the colorimeter
AMBIENT TEMP
Air temperature inside the electronics area
SAMPLE TEMP
Sample pre-heater block temperature (typically 45 °C to 55 °C (113 °F to 131 °F) but
could reach 58 °C (136.4 °F))
AIR PRESS
Reagent air pressure in the reagent bottles (ideally 3.95 to 4.10 psi)
LED DUTY CYCLE
Depends on condition of colorimeter cell and age of analyzer (typically 7,200 to
40,000 counts)
HEATER DUTY CYCLE Percentage of time the colorimeter heater is on to maintain a constant 50 °C (122 °F)
SAMPLE FLOW
Approximate sample flow into the colorimeter, measured during the flush cycle
SAMPLE PRESS 1
Sample pressure before the pre-heater block (ideally 2 to 4.5 psi depending on
incoming sample pressure)
SAMPLE PRESS 2
Sample pressure after the sample pre-heater, which is used to calculate the sample
flow. The sample pressure is almost zero when the flush is set to off and is
approximately 0.2 psi when the flush is on (depends on inlet pressure and flow).
REAGENT 1
Level of reagent remaining
REAGENT 2
Level of reagent remaining
REAGENT 3
Level of reagent remaining
STD SOLUTION
Level of standard solution remaining
FAN SPEED
The speed of the ventilation fan.
LEAK COUNTS
Indication of possible fluid leaks (range 0 to 1023). A count of over 511 indicates a fluid
leak
Table 3 Measurement data
Element
Definition
LAST MEAS CHANNEL Last channel measured.
LAST MEAS TIME
Time of last measurement.
English
15
Summary of Contents for 5500sc PO4 3-HR
Page 208: ...208 216 211 221 213 SD 224 213 224 214 1 1 1 6 2 7 3 8 4 9 5 Enter 2 3 208...
Page 209: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 4 0 100 1 209...
Page 210: ...45 75 10 45 50 75 10 50 2 6 1 2 3 ppb 48 SD 4 SD 210...
Page 211: ...4 SD 1 3 2 2 4 SD 1 1 1 2 3 4 211 4 1 2 3 4 2 211...
Page 212: ...5 6 7 0 14 bar 2 psi 0 28 bar 4 psi 55 mL 8 ENTER 9 10 224 216 216 1 enter enter 1 224 212...
Page 213: ...MSDS SDS 1 2 1 100 1 100 1 100 100 100 100 1 250 500 mL 2 213...
Page 214: ...3 4 5 5 5 1 2 3 4 211 1 2 3 9 10 240 15 ppb ppm mg L g L 1 5 1 16 10 214...
Page 215: ...10 216 223 216 4 20 mA 216 Modbus Profibus HART HACH55 215...
Page 218: ...PID 5 50 50 6 1 2 3 219 216 218...
Page 219: ...1 2 3 1 0 0 9 1 0 1 1 1 0 0 5 1 0 1 5 5 4 0 0 5 3 5 4 0 5 0 300 5 0 300 5 4 10 219...
Page 220: ...5 5 5 0 0 0 0 6 5 30 7 8 1 220...
Page 221: ...2 3 4 5 6 1 1 2 18 000 18 000 1 2 2 3 221...
Page 224: ...SD SD SD SD 2 GB 1 SD 4 211 2 SD SD SD 3 SD SD SD SD 1 SD 2 SD SD SD 3 4 SD 5 1 224...
Page 225: ...1 225 225 4 20 mA 1 2 3 ppb 4 1 2 ENTER 3 4 225...
Page 279: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 PROGNOSYS 4 PROGNOSYS PROGNOSYS 0 100 1 279...
Page 280: ...45 75 10 45 50 75 10 50 2 1 2 home 3 48 SD 4 SD 280...
Page 281: ...4 SD 1 3 2 2 4 SD 1 1 1 2 3 4 281 4 1 2 3 4 5 6 2 281...
Page 282: ...7 0 14 2 0 28 4 55 8 enter 9 10 294 286 286 1 enter enter 1 294 MSDS SDS 282...
Page 283: ...1 menu 2 1 100 1 100 1 100 100 100 100 1 250 500 2 3 4 5 5 5 283...
Page 284: ...1 2 3 4 281 1 2 menu 3 9 10 240 15 ppb ppm 1 5 1 16 10 10 284...
Page 285: ...285 292 286 4 20 286 Modbus Profibus HART HACH55 1 menu 2 285...
Page 287: ...4 20 1 menu 4 20 2 3 4 5 50 50 6 287...
Page 288: ...1 menu 2 3 288 286 1 menu 2 3 1 0 0 9 1 0 1 1 1 0 0 5 1 0 1 5 5 288...
Page 289: ...4 0 0 5 3 5 4 0 5 0 300 5 0 300 5 4 10 5 5 5 ONMAX 0 OFFMAX 0 ONMIN 0 min OFFMIN 0 min 6 289...
Page 290: ...5 30 7 8 1 2 3 4 5 6 1 1 menu 2 290...
Page 292: ...2 0 1023 511 3 A D REF A D A D 6 6 6 1 2 3 1 2 292...
Page 294: ...1 SD 2 SD SD SD 3 4 SD 5 1 1 cal 295 294 4 20 1 cal 2 3 ppb 4 294...
Page 295: ...1 cal 2 enter 3 4 295...
Page 370: ...4 SD 1 3 LED 2 2 4 SD 1 1 1 2 3 4 370 4 1 LANGUAGE 2 DATE FORMAT 3 DATE TIME 2 370...
Page 386: ...1 cal START MANUAL CAL 2 enter 3 YES NO 4 386...
Page 404: ......
Page 405: ......