Option
Description
CALIBRATION DATA
Displays data from the last calibration and the due date and time of the next
scheduled calibration.
OVERRIDE CALIBRATION
Enter new slope and/or zero value (reagent blank). When OVERRIDE
CALIBRATION is selected, the automatic calibration is disabled for the selected
option.
OUTPUT CALIBRATION
Select a 4-20 mA output and enter the output values to be sent.
RESET DEFAULT CAL
Reset calibration data to default values and disables automatic calibration. On
completion, perform a new calibration.
Schedule automatic calibrations
1.
Push
cal
, then select SET AUTO CALIBRATION.
2.
Select ACTIVATE AUTO CAL>YES.
3.
Select STD SOLUTION and enter the standard value in ppb.
4.
Select a schedule option for calibration.
Option
Description
TIME BASE
Sets the interval between calibrations. Options: DAYS or HOURS.
WEEK DAY
Selects the day or days of the week for calibration when TIME BASE is set to DAY.
TIME
Sets the time of day for calibration when the TIME BASE is set to DAY.
SET INTERVAL
Sets the interval between automatic calibrations in hours when the TIME BASE is set to
HOURS .
Start a calibration manually
1.
Push
cal
, then select START MANUAL CAL.
2.
The type of calibration is displayed. Push
enter
to start the calibration.
3.
The measurement status is displayed. Select YES to interrupt the current measurement cycle and
start the calibration immediately. Select NO to wait until the current measurement cycle has
completed before the calibration starts.
4.
Complete the instructions on the display.
18
English
Summary of Contents for 500sc SiO2
Page 1: ...DOC023 97 80267 5500sc SiO2 07 2016 Edition 2 Operations Fonctionnement Operaciones Opera es...
Page 2: ...English 3 Fran ais 19 Espa ol 36 Portugu s 54 71 86 103 119 2...
Page 89: ...4 SD 1 3 LED 2 2 4 SD 1 1 1 2 3 4 89 4 1 LANGUAGE 2 DATE FORMAT 3 DATE TIME 4 YES 5 6 7 2 89...
Page 134: ......
Page 135: ......