Specifications
Specifications subject to change without notice.
Application
For water samples. Do not use the application for concentrated acids,
bases or any type of volatile sample.
Output temperature precision
±1.5 ºC (2.7 ºF) under reference conditions
1
Display resolution
1 ºC (1 ºF)
Set point range
20 to 50 °C (68 to 122 °F) Refer to operating charts for more information.
Flow rate range
50-300 mL/min; Sample Heater internal volume is 39 mL
Power requirements for single
phase (with grounded neutral) only
120 Vac, 50/60 Hz, 11A
230 Vac, 50/60 Hz, 5.5A
Installation category
II
Inlet pressure
689.5 kPa (100 psi) max. internal pressure. This rating is for the Sample
Heater and does not affect the Analyzer requirements.
Sample fittings
Two ¼" OD Tubing Compression fittings
Weight
Net—4.98 kg (11 lb)
Shipping—6.35 kg (14 lb)
Wetted parts
316 Stainless Steel
Ambient operating temperature
0 to 50 °C (32 to 122 °F)
Pollution Degree
2
Protection Class
I
Maximum Altitude
2000 m (6560 ft)
Enclosure Rating
NEMA 4X, IP 65
Compliance Information
Listed for use in general locations to UL and CSA Standards by ETL. CE
certified.
Storage Temperature
-20 to 70 ºC (-4 to 158 ºF) with fluids drained.
Humidity
Maximum relative humidity 80% for temperatures up to 31 ºC decreasing
linearly to 50% R.H. at 40 ºC.
1
Within proportional control range (see charts) and input temperature ±5 ºC, flow rate ±10%, 115 VAC, 60 Hz
±2 Hz and ambient temperature ± ºC. Instrument tuned at operating conditions (Autotune 3). Cal offset
adjusted to yield the set temperature at the point of use.
Sample heater operating range
shows in graph form the sample heater operating range in terms of sample output
shows in graph form the sample heater operating range in terms
of sample output temperature and line AC voltage.