29
Troubleshooting
7.3 Event Codes
Event codes are not displayed on controller and must be downloaded from the event log to
be viewed. Troubleshooting actions are provided in
Table 5 on page 28
.
High Readings
Deionized water turbidity is greater than
0.5 NTU
Clean the instrument.
Access Calibration History for turbidity value of
ultra-filtered water.
Verify the flow is between 200–750 mL/min.
Recalibrate the instrument.
Table 6 Additional Malfunctions Not Recorded in the Event Log (continued)
Symptom
Possible Cause
Corrective Action
Table 7 Event Log List
Event
Event #
Data1
Data2
Data3
Bubble Reject Change
0
0 = OFF
1= ON
—
—
Filter Size Change
1
0 = No avg, 1 = 6 sec
2 = 30 sec, 3 = 60 sec
4 = 90 sec
—
—
Data Log Interval Change
2
0 = 30 sec, 1 = 1min
2 = 5 min, 3 = 10 min
4 = 15 min
—
—
Power On
3
1
—
—
Calibration
4
Cal Gain
Initials
—
Verification
5
Expected Value
Meas Value
Initials
Dark Reading Warning
6
Measured Value
—
—
Temperature Warning
7
Temperature Value
—
—
Voltage Warning
8
16 = +5V high
32 = +5V low
64 = V in high
128 = V in low
—
—
Lamp Warning
9
1 = lamp V high
2 = lamp V low
4 = lamp I high
8 = lamp I low
—
—
A/D Fail
10
1
—
—
Lamp Fail
11
2
—
—
Flash Fail
12
4
—
—
Output Mode Change
13
0 = Active, 1 = Hold
2 = Transfer, 4 = Normal
—
—
Summary of Contents for 1720E
Page 1: ...Catalog Number 6010018 1720E Low Range Turbidimeter USER MANUAL 04 2016 Edition 8...
Page 2: ......
Page 12: ...10 Installation Figure 4 1720E Dimensions...
Page 16: ...14 System Startup...
Page 24: ...22 Operation...
Page 32: ...30 Troubleshooting...
Page 38: ...36 Modbus Register Information...
Page 40: ...38 Index...
Page 41: ......