Transmitter 9125 – Conductivity / resistivity measurement
Page 40
TEMPERATURE COMPENSATION
MEASURE :
- No
- Pt 100
- Pt1000
Choice of a temperature
measurement with or without
Pt100 / Pt1000.
TYPE :
- No
- Auto.
- Manual
Choice between no temperature
compensation or an automatic or a
manual temperature compensation
mode.
TEMP. :
XX
Possibility to enter the sample
temperature in a manual
compensation.
TREF :
XX
Possibility to enter the reference
temperature.
COMP. :
- Coef.
- HCl
- NaCl
Possibility to choose the
temperature compensation.
COEF :
XX
Possibility to enter the coefficient
value.
TDS coefficient adjusting
TDS coefficient adjusting is directly in the MEASURE
menu.
Utilize the
left
function touch to choose the TDS
adjusting.
TDS
: Choose the TDS coefficient which is necessary
to calculate the solution concentration.
Utilize the
right
function touch to choose a value
between 0 and 5.
MEASURE :
TYPE : Manual
TEMP. : 15°C
TREF. : 025.0°C
COMP. : Coeff.
COEF. : 2.0%
TEMP. COMP.
Select
Pt100
PROBE
TEMP. COMP. :
TDS :
0.50
MEASURE
Select
0.50