Figure 2 Product components
1
Turbidity sensor with 1.6 m cable (3 ft 3
⅜
in.)
6
Screws, bracket to sensor (4x), 4 x 16 mm
2
Vial removal tool
7
Screws for tubing clips (4x), 2.2 x 6 mm
3
Maintenance rack
1
8
Tubing clips (4x)
4
Vial wiper
9
Bracket for tubing attachments and optional
accessory
5
Desiccant cartridge
10
High-temperature screws (3x) for (40 to 70 ºC
(104 to 158 ºF))
1
Not supplied with the TU5300 sc EPA model.
Installation
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Installation guidelines
N O T I C E
Make sure that a floor drain is available near the instrument. Examine the instrument daily for leaks.
This instrument is rated for an altitude of 3100 m (10,710 ft) maximum. Use of this instrument at an
altitude higher than 3100 m can slightly increase the potential for the insulation to breakdown, which
can result in an electric shock hazard. The manufacturer recommends that users with concerns
contact technical support.
• Install the sensor upright and level on a flat, vertical surface.
• Mounting hardware is supplied by the user.
• Refer to the documentation supplied with the accessories for installation of the optional units.
• Refer to
on page 10 for the minimum clearances.
Installation overview
shows the basic installation overview. The sensor can be used with optional accessories.
Refer to the documentation supplied with the accessory. Refer to the expanded manual on the
manufacturer's website for a full installation overview.
10
English