HACH LANGE TU5400 sc Basic User Manual Download Page 4

Specification

Details

Calibration options

StablCal

®

: at 20  NTU from 0 to 40  NTU; at 20  NTU and 600  NTU for full

range
User prepared formazin: at 20  NTU from 0 to 40  NTU; at 20  NTU and 600 
NTU for full range
Custom calibration for up to 6-point calibrations

Verification options

Liquid: StablCal

®

, formazin (0.1 to 40 NTU)

Dry: factory-set values of 1.0 ± 25% NTU

Verification (Link2SC software)

Verfication of the measurement value to compare values to a laboratory
instrument via Link2sc software

Dimensions (W x D x H)

268 x 249 x 190 mm (10.6 x 9.8 x 7.5 in.)

Enclosure

Rating: NEMA/IP65 for functional units
Material: ASA Luran S 777K / RAL7000, TPE RESIN Elastocon

®

 STK40,

Thermoplastic Elastomer TPS-SEBS (60 Shore)

Display resolution

0.0001  NTU (0.0001 to 0.9999/1.000 to 9.999/10.00 to 99.99/100.0 to
700 NTU)
Default=TU5300sc: 0.001 NTU, TU5400sc: 0.0001 NTU

Sensor cable length

1.6 m (5.25 ft), extendable up to 50 m (164 ft) for instrument without
accessories

Fittings

Sample quick connector: ¼-in. for ¼-in. tubing

Tubing

Materials: Polyethylene, polyamide or polyurethane tubing, calibrated outside
diameter ¼ in. +0.001 / –0.004 in. or +0.03 mm / –0.1 mm

Weight

???

Installation environment

Indoor or outdoor with cover

Mounting

Wall

Protection class

III

Pollution degree/installation
category

2, II

Power requirements

12 VDC, 14 VA from the sc controller. Refer to the controller documentation
for the controller power requirements.

Operating temperature

0 to 50 °C (32 to 122 °F)

Storage temperature

–40 to 60 °C (–40 to 140 °F)

Operating humidity

Relative humidity: 5 to 95% at different temperatures, non-condensing

Certifications

CE compliant; US FDA accession number: 1420493-xxx. This product
complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with
Laser Notice No. 50.

Warranty

1 year (EU: 2 years)

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.

4

   

English

-BETA DRAFT-

BETA

Summary of Contents for TU5400 sc

Page 1: ...DOC023 52 90501 TU5400 sc TU5300 sc 09 2014 Edition 1 Basic User Manual BETA DRAFT B E T A...

Page 2: ...BETA DRAFT B E T A...

Page 3: ...NTU from 0 to 40 NTU 10 of reading from 40 to 700 NTU based on formazin primary standard Linearity Better than 1 for 0 to 40 NTU on formazin at 25 C 77 F Allows for accurate calibration at 20 NTU Repe...

Page 4: ...ubing Tubing Materials Polyethylene polyamide or polyurethane tubing calibrated outside diameter in 0 001 0 004 in or 0 03 mm 0 1 mm Weight Installation environment Indoor or outdoor with cover Mounti...

Page 5: ...death or serious injury W A R N I N G Indicates a potentially or imminently hazardous situation which if not avoided could result in death or serious injury C A U T I O N Indicates a potentially hazar...

Page 6: ...er radiation when the instrument is defective and when the instrument lid is open This product complies with EN 61010 1 Safety Requirements for Electrical Equipment for Measurement Control and Laborat...

Page 7: ...shock The instrument can cause interference in immediate proximity to televisions radios and computers The warranty does not cover improper use or wear FCC conformance for RFID This instrument may con...

Page 8: ...overview D A N G E R Chemical or biological hazards If this instrument is used to monitor a treatment process and or chemical feed system for which there are regulatory limits and monitoring requireme...

Page 9: ...F for the vial access cover1 11 Process head closed 5 Vial access cover 12 Cable channels 6 Inlet 13 Extension connector for more accessories 7 Outlet 14 Sensor cable 1 Use a screwdriver to change th...

Page 10: ...instrument is rated for an altitude of 3100 m 10 710 ft maximum Use of this instrument at an altitude higher than 3100 m can slightly increase the potential for the insulation to breakdown which can...

Page 11: ...tlet 3 Controller 6 Sensor Wall mount Install with a bracket D A N G E R Electrocution hazard Remove power from the instrument before this procedure is started To install the sensor on a wall with the...

Page 12: ...12 English BETA DRAFT B E T A...

Page 13: ...ted steps that follow Make sure to remove the thin plastic film from the mounting holes on the back of the instrument Install the maintenance rack Refer to the illustrated steps that follow to install...

Page 14: ...sc controller Connect the sensor to the quick connect fitting of an sc controller Refer to Figure 5 The sc controller can be on or off Keep the connector cap for later use Figure 5 Quick connect fitt...

Page 15: ...ntil the instrument is ready for operation again Yellow stable Warning for a time out e g during a calibration Condensation warning auto verification Contamination warning auto verification Yellow fla...

Page 16: ...rd The sample line contains water under high water pressure that can burn skin if hot Qualified personnel must remove the water pressure and wear personal protective equipment during this procedure N...

Page 17: ...n and more Refer to the expanded manual on the manufacturer s website DIAG TEST Gives diagnostics and maintenance options Refer to Maintenance on page 17 Calibration W A R N I N G Chemical exposure ha...

Page 18: ...may increase the frequency of some tasks Table 3 Maintenance schedule Task 3 weeks 3 months 1 year Refer to Clean the vial with the wiper on page 20 settled water 1 X Refer to Clean the vial with the...

Page 19: ...aced VIAL Shows the operational time of the vial Resets the counter when the vial is replaced in Replace the vial on page 20 CLEANING Shows the last cleaning date Resets the counter when the vial is c...

Page 20: ...ded process in MAINTENANCE VIAL CLEANING Replace the vial N O T I C E Do not touch the glass of the vial Contamination on the glass can cause measurement errors To protect the glass from contamination...

Page 21: ...English 21 BETA DRAFT B E T A...

Page 22: ...fer to Plumb the sensor on page 16 to replace the tubing Clean the vial compartment Clean the vial compartment only when the compartment is contaminated Table 4 gives options on how to clean the vial...

Page 23: ...BETA DRAFT B E T A...

Page 24: ...w hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Compois 1222...

Reviews: