background image

 

HACH-LANGE, Willstätter Str. 11, D-40549 Düsseldorf, Germany 

Telephone: +49 (0)211/5288-0; FAX: +49 (0)211/5288-143 

sc1000 Software Updates from SD Card 

Table of Contents 

 

Table of Contents ___________________________________________________________________  1 

The sc1000 SD Card Connector/Slot _____________________________________________________  1 

Software Update of sc1000 Display module (LXV402) _______________________________________  4 

 

The sc1000 SD Card Connector/Slot 

 
The sc1000 SD Card Connector is used to connect  

 

MMC cards up to 1GB, 

 

SD cards up to 1GB, 

 

SD cards of 2GB 

(1)

 

SD cards compliantwith the SDHC standard (4 – 32GB) 

(1)

 
Note (1): 

Cards with capacities of ≥ 2GB can only be connected, when sc1000 software v3.20 or higher 

is running

 

 

Functionality of the SD Card Connector/Slot 

 

Update the sc1000 system: All components of the sc1000 controller system and all connected 
sc sensors can get software updates by uploading new software versions from the card to the 
instruments. 

 

Saving and restoring configuration and settings of devices, which are connected to the sc1000 
controller. Device data is stored to the card. This data can later be sent back to the devices, to 
restore the saved status.  
Note: Restoring settings and configuration does not restore the content of data and event log. 

 

Storing diagnostics data for a detailed analysis of error conditions by Service. 
Getting (one time action) and logging (continuous action) of Data and Event log files from the 
connected devices. The files are in <.csv> -format and can be imported to Excel to create 
curves or to explore the data in detail. 

 

How to get access to the sc1000 using the SD Card Connector/Slot 

 
Required hardware:  

 

Storage card with a size of 1GB or less for sc1000 Display software version 3.12 or lower. 

 

MMC Adapter (part no. YAB 902) for old 

(2)

 sc1000 Display Modules. This package already 

contains a 1GB SD card. 

 
 

 
 
 

Summary of Contents for sc1000

Page 1: ... controller system and all connected sc sensors can get software updates by uploading new software versions from the card to the instruments Saving and restoring configuration and settings of devices which are connected to the sc1000 controller Device data is stored to the card This data can later be sent back to the devices to restore the saved status Note Restoring settings and configuration doe...

Page 2: ...MC Adapter part no YAB 902 for connecting storage cards this applies for all types of cards MMC SD and SDHC See figure2 figure2 Connecting storage cards to old sc1000 controllers Connect the Storage Card to the sc1000 Display contact stripes of the card to the bottom side When an MMC adapter is used figure2 first insert the card to the adapter and then insert the adapter into the card slot of the ...

Page 3: ...SETUP screen will not display the STORAGE CARD option Note After a Storage card is connected it can take from 10 seconds up to several minutes before the STORAGE CARD option will be visible in the menu The time depends on card size and on the number of files that are stored on the card To minimize wait time it is recommended to use an SD card which has been cleared or even formatted before using i...

Page 4: ... Laptop Create a folder named update immediately at the root of the Storage Card Copy Software Updates to this folder figure5 Software Updates on Storage card Update Software from Storage Card Ensure that all steps described in How to get access to the sc1000 using the SD Card Connector Slot have been completed Follow the menu tree to access contents of the SD card MENU SC1000 SETUP STORAGE CARD S...

Page 5: ...ning software a notification is shown figure7 sc1000 Software Update menu without recent files In menu figure6 sc1000 Software Update menu press the OK button to get to the final confirmation which starts the update process figure8 figure8 Confirmation menu before Updates start Note Confirming stops the normal operation of the sc1000 controller From now only the software updates for the sc1000 con...

Page 6: ...is state because the unit can be permanently damaged Note Starting from sc1000 Software v3 20 size indicators show the progress during updates This gives users the possibility to estimate the remaining update time Look at the end of the progress messages by pressing the Down button and verify that the software has finished When no errors occurred the following lines are shown Update s complete Ple...

Page 7: ...ätter Str 11 D 40549 Düsseldorf Germany Telephone 49 0 211 5288 0 FAX 49 0 211 5288 143 figure10 End of Software Update The Software Update is finished Repower the sc1000 display module Note Disconnecting the sc1000 controller during Software Updates will damage the unit ...

Reviews: