© HAAG-STREIT AG, 3098 Koeniz, Switzerland – HS Doc. No. 1500.7220551.04020 – 2016 – 08
3.4 Cabling of the CM03 and the RM02
(diagram)
WARNING!
•
Only use the supplied USB 3.0 cable (3m long, HS art. no. 1022373)
for the connection to the PC.
•
With lengths of over 3m, the cable with active signal amplification
(HS art. no. 1022441) must be used.
•
Only use medically approved PCs or operate via a medically ap‑
proved isolating transformer.
•
Auxiliary units on the PC (e.g. printer, monitor) must be operated
through an isolating transformer.
•
Ethernet may only be used through a galvanic isolation in accor‑
dance with EN 60601‑1.
•
The power supply unit's mains connector must be accessible in order
to allow for disconnection from the electric mains at any time!
25. Power grid
26. Med. approved isolating transformer
27. Instrument table (IT)
28. Med. approved power supply
29. Printer
30. Screen
31. Personal computer
32. Local network
33. Galvanic isolation (EN 60601‑1)
34. Headrest
35. Cable headrest / LED illumination /
RM02
36. Release module RM02
37. LED illumination
38. Camera module CM03
25
26
27
29
30
31
34
36
37
38
32
33
35
28
NOTE!
To ensure that the system works correctly, HAAG‑STREIT recom‑
mends not using laptops and using a high‑quality desktop computer
instead.
39. Fastening screw
40. Two‑pole connection plug
If the middle LED lights up red during
operation, the two‑pole connection
plug (40) is connected incorrectly.
•
Disconnect the device from the pow‑
er grid.
•
Remove the cover on the upper part
of the illumination facility by loosen‑
ing the fastening screw (39).
180°
38
39
1
39
40
1
•
Turn the two‑pole connection plug (40) 180°.
•
Fix the cover on the upper part of the illumination facility with the fastening screw
(39).
•
Connect the device to the power grid again.
WARNING!
•
Keep magnet-sensitive storage media (e.g. credit cards) away from
the magnets on the release module RM02!
•
Only external medical power supplies approved by Haag‑Streit that
fulfill EN 60601-1 may be used.
3.4.1 Step-by-step cabling
•
Place the release module RM02 over the slit lamp's cross slide. Four magnets are
used for fixing.
NOTE!
With BQ 900 slit lamps with a date of manufacture before 1998, the
cover plate is fixed on the crosstree carriage with screws. The two
screws at the back must be removed before the RM02 is mounted.
•
Insert the camera cable connector plug (50) in the socket (59).
•
Press the camera cable into the cable holders (49).
•
Computer cable USB 3.0 (47), power supply cable (46) and cable headrest / LED
illumination / RM02 (45) must be fed through in the braided sleeving (44).
•
Insert the connector plug of the power supply cable (46) in the socket (59).
01-IFU_IM900-7220551-04020_eng.indd 8