WARNING!
The software must be installed by trained personnel in accordance with
the separate installation instructions. For further information, please
contact your Haag-Streit representative.
NOTE!
• The instructions for use is integrated into the software.
• Error messages are described in the software instructions for use.
8 Maintenance
The Eyestar 900 is practically maintenance-free and requires only minimum care to
work for as long as possible to your complete satisfaction. However, we recommend
instructing a service engineer to inspect the biometer periodically. Haag-Streit or
your local representative will be happy to provide further information.
8.1 Cleaning
The cleaning intervals for cleaning the complete device shall be performed within
reasonable time (e.g. once a week). Cleaning of applied parts shall be performed
prior to the measurement procedure with a new patient.
Please use only following tools for cleaning operations:
• Cleaning liquid: Isopropanol 2 (cleaning liquid with maximum 70 % alcohol)
• Cover towel: Commercially available towel
• Display towel: Commercially available towel for cleaning screens
• HS Lenspen (see appendix)
In particular, attention has to be paid to the following points:
• Only use cleaning liquid evenly spread on the cover towel on following device
parts:
• Patient interface chin rest shell including its supporting parts
• Patient interface forehead rest band including its supporting parts
• Patient interface handles including their supporting parts
• Device cover (casing)
• User interface display panel
• User interface display panel surroundings and stand
The device patient interface cover glasses shall only be cleaned with the HS
Lenspen.
WARNING!
Only perform maintenance on a powered off device.
We suggest following cleaning procedure:
• Put cleaning liquid evenly spread on the cover towel
• Clean the patient interface forehead rest band, chin rest shell and both handles
including their supporting parts with the cover towel
• Put cleaning liquid evenly spread on the cover towel
• Clean the complete device cover with the cover towel
• Put cleaning liquid evenly spread on the cover towel
• Clean the display panel surroundings and stand with the cover towel
• Clean the display panel (control panel) with the display towel
• Clean the patient interface cover glasses with the HS Lenspen with the felt and
the brush
• Protect the device from dust with the dedicated accessory dust cover when the
device is not in use.
WARNING!
The device must not be switched on when covered! (Heat build-up, fire
hazard).