Slit image rotatability

± 90°

Swiveling of the slit illumination to 

the microscope axis

Horizontal ± 90°, vertical 0 – 20°

Filters

Blue, red removal (green), gray (10%).

NOTE!

(Further information available in the LED illumination LI 900 instruc-

tions for use)

7.2  Stereo microscope 

Stereo angle:

13°

Magnification changer:

6.3 × / 10 × / 16 × / 25 × / 40 ×

Ocular magnification:

12.5 ×

Range of adjusting eye-pieces:

+7 to −7 diopters

Pupil distance:

52 – 78 mm

Total magnification:

6.3 × / 10 × / 16 × / 25 × / 40 ×

Object field ø in mm:

32.0 / 20.0 / 12.7 / 8.0 / 5.1

 

7.3  Instrument base 

Operation:

Single-handed operation of control lever in 3 dimensions

Adjustment of instru-

ment base:

100 mm (length) 

100 mm (side) 

30 mm (height)

7.4 Dimensions

Weight:

12.7 kg (without power supply, head rest and options)

Dimensions L × W × H: 305 × 332 × 700 mm
Packaging L × W × H:  420 × 510 × 780 mm

 

8. Maintenance

WARNING!

 

Do not modify this equipment without authorization of the manufacturer.

Installation and repairs may only be performed by trained specialists. 

Contact your Haag-Streit representative for installation, repairs and 

modification work on the system. The contact details are available at 

www.haag-streit.com.

Only original Haag-Streit replacement parts may be used.

 

 

The LED illumination can be operated maintenance-free for its entire service life.

8.1  Device inspection 

In order to correctly check the slit lamp, proceed as follows:

Insert test rod into the radial movement bearing, whilst at the same time aligning

   the surface to the microscope at a right angle.

Set slit length to 8 or 14 mm.

Set llumination strength to 50%.

Set magnification to max. in the microscope.

Set the eyepieces in such a way that the test rod is in sharp focus. Turn the 

  eyepiece from the (+) to the (-) side.

The structure of the test rod must be in sharp focus in all magnifications.

Close slit edges to approx. 0.5 mm. The borders must be in sharp focus here.

Completely open slit edges and turn the test rod by 45°, the sharp area must 

  be in the centre of the test rod.

8.2 Repair

To guarantee a long service life, the device must be cleaned weekly as described 

and protected with the dust cover when not in use. We recommend having the de-

vice inspected once a year by an authorized service technician.

In order to guarantee optimal operational integrity, the device must undergo periodic 

maintenance. The length of the maintenance interval depends on use, but may not 

exceed 3 years.

8.3  Cleaning and disinfection

The Haag-Streit slit lamps and their accessories can, if required, be carefully wiped 

down with ready-for-use disposable 70% ethanol disinfectant wipes. Surface-friend

-

ly disinfectants (containing aldehyde or aldehyde-free) are also permitted, such as 

Kohrsolin FF.

WARNING!

• 

The preparation instructions provided do not apply to tonometer mea-

suring prisms!

• 

Tonometer measuring prisms must be prepared in accordance with a 

different manual

• 

Do not use sprays

• 

Observe the manufacturer's safety instructions

• 

Do not use any cloths that drip.

01-IFU_BQ900-7220589-04200_eng.indd   10

07.06.2019   11:30:45

Summary of Contents for BQ 900

Page 1: ...ANO ESPA OL NEDERLANDS PORTUGU S SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7220589 04200 2019 05 INSTRUCTIONS FOR USE Slit Lamp BQ 900 20 Edition 2019 05 01 IFU_BQ900 7220589 04200...

Page 2: ...eases or trauma which affect the structural properties of the eye Contraindication There is no absolute contraindication for tests with this device Appropriate profes sional judgement and caution are...

Page 3: ...ssioning 7 4 1 Switching on the appliance 7 5 Operation 7 5 1 Setting the eyepieces 7 5 2 Preparing the patient 8 5 3 Operating the instrument 8 5 4 Setting the filters diaphragms 9 5 5 Fixation star...

Page 4: ...t handling or damage If this is the case notify the transport company that has delivered the goods to you Unpack the equipment together with a represen tative of the transport company Make a report of...

Page 5: ...es cribed in the documents distributed with the product The product must be treated as described in the Safety chapter Improper hand ling can damage the product This would void all guarantee claims Co...

Page 6: ...ssories pin 24 Quick release fastener for accessories 25 Stereo microscope with eyepieces 26 Eyepieces 27 Breath shield 28 Mounting screw for breath shield 29 Thread for fixing right hand side the ton...

Page 7: ...springs 3 7 Regulating the clearance of the slit width controls The small screw in the center of the right control knob A allows you to regulate the friction of the turning movement of these adjusting...

Page 8: ...verified with the naked eye Fine tuning is performed by tilting the control lever while observing via the stereo microscope 25 The slit width is set left or right with the rotating knob 19 as is the...

Page 9: ...re moval filter 5 6 Microscope and eyepiece 40 Front lens 41 Rotary knob displaying the set magnification 42 5 level magnification changer Galilei system 43 Knurled ring bayonet joint 44 Binocular tub...

Page 10: ...it lamp proceed as follows Insert test rod into the radial movement bearing whilst at the same time aligning the surface to the microscope at a right angle Set slit length to 8 or 14 mm Set lluminatio...

Page 11: ...le BM900 To be used in combination with 7220737 7020028 Stereo variator 7200109 Second observer tube long 7220146 Second observer tube short 7220147 Eyepiece 12 5 with crosshair reticule 7220454 Eyepi...

Page 12: ...ss I FDA Class II D Disposal Electrical and electronic devices must be disposed of separately from household waste This appliance was made available for sale after the 13th August 2005 For correct dis...

Page 13: ...ce to interfe rence Third party devices may only be connected in compliance with the EN 60601 1 standard F 2 Emitted interference standard table 1 The information is based on the requirements of EN 60...

Page 14: ...spital environment Surge EN 61000 4 5 1 kV for symmetrical voltages 2 kV for asymmetrical voltages 1 kV for symmetrical voltages 2 kV for asymmetrical voltages Mains power quality should be that of a...

Page 15: ...paration distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference...

Page 16: ...output of the transmitter W Safe distance according to transmission frequency m 150 kHz 80 MHz D 0 35 80 MHz 800 MHz D 0 7 800 MHz 2 7 GHz D 1 4 0 01 0 1 1 10 100 For transmitters with a nominal outpu...