Haag-Streit AT 870 Instructions For Use Manual Download Page 9

9

DEUTSCH

ENGLISH 

FRANÇAIS

ITALIANO

ESPAÑOL

NEDERLANDS

PORTUGUÊS

SVENSKA

DEUTSCH

ENGLISH

FRANÇAIS

ITALIANO

ESPAÑOL

NEDERLANDS

PORTUGUÊS

SVENSKA

© HAAG‑STREIT AG, 3098 Koeniz, Switzerland – HS‑Doc. no. 1500.7006000.04260 – 26. Edition / 2020 – 03

NOTE! 

A smooth operation can only be guaranteed with the use of original 

Haag‑Streit measuring prisms and Tonosafe. 

1.  Anaesthetise each eye

2.  Apply Fluorescein into the eye to be examined 

3.  The correct eye height of the patient can be set via the chin rest.

5.4  Patient instructions

1. 

Press the head firmly against the chin and forehead rests.

2. 

The patient must look straight forward. If necessary, the small fixing light can be 

used to steady the eyes.

3.  It is recommended that the patient be repeatedly reminded to keep his eyes 

open wide during the examination. It might be necessary for the examiner to 

keep the eye open by splaying the eyelids with thumb and index finger. 

4.  This must be done without applying any pressure to the eye. 

5.5  Preparation of slit lamp and tonometer

For all Haag-Streit slit lamps and tonometers

1.  Before the examination, the eyepieces must be adjusted correctly to the exam‑

iner.

2. 

Set the magnification to 10×.

3.  Adjust the illumination to medium intensity.

4. 

Position the blue filter in the beam path of the slit lamp's illumination apparatus 

and open the slit diaphragm fully.

5.  Insert the disinfected measuring prism into the (0° position) holder on the sensor 

arm. For Tonosafe, see the separate instructions for use.

6.  Snap the sensor arm into place so that the axes of the measuring prism and of 

the microscope align.

7. 

Set the measuring drum to the first calibration line.

AT 900 model R and AT 900 model BQ

1.  Swing the illumination apparatus to the left.

2.  Release the tonometer from the dwell position to the right of the microscope, 

and swing it forward until it locks in the measuring position.

3.  From the left, bring the illumination apparatus into contact with the tonometer 

bearer arm. This is the only illumination position in which both the patient’s left 

and right eyes can be easily examined (no 60° position). This arrangement facili‑

tates the splaying of the patient’s eyelids, should this be necessary for measure‑

ment. The illumination of the applanated surface through the measuring prism is 

practically reflection-free.

Observation:

   with AT 900 model R in the left eyepiece

 

 

with AT 900 model BQ in the right eyepiece

AT 900 model T

1.  For an examination through the tonometer’s left or right eyepiece, the angle 

between the illumination instrument and the microscope should be approx. 60° 

so that the image is bright and reflection-free. Alternatively: lighting through the 

measuring prism at approx. 10°.

AT 870

1.  Swing the sensor arm with the measuring prism into the beam path of the micro‑

scope and the illumination.

2.  The angle between the illumination equipment and the microscope should be 

approximately 60° so that the image appears bright and free of reflections. Alter

natively: lighting through the measuring prism at approx. 10°.

3. 

Set the measuring drum to the first calibration line.

4.  Set a mid‑level illumination intensity.

WARNING! 

Measurements through observation with the wrong ocular can lead to 

false results.

5.6  Measuring correctly

NOTE!

Make sure that the measuring arm of the tonometer can move freely and 

is not hindered by the patient's eyelashes, beard, etc.

1.  Immediately before taking measurements, the patient should close his eyes 

briefly so that the cornea becomes sufficiently moistened with fluorescein-im

pregnated tear fluid.

2.  By moving the slit lamp, the measuring prism comes into contact with the centre 

of the cornea over the pupillary area.

3.  During contact, the cornea's limbus takes on a bluish glow. This glow can be 

best observed with the naked eye from the opposite side of the illumination ap‑

paratus.

4.  When the limbus glows, stop moving the slit lamp immediately.

5.  After contact is made, viewing is conducted through the microscope. The uni‑

form pulsation of the two semi circular fluorescein bands, which could be of dif

ferent sizes in drum setting 1 depending upon inter‑ocular pressure, shows that 

the tonometer is in the right measuring position.

6.  Any necessary corrections are done using the slit lamp control lever, until the 

1500_7006000_04260_Gebrauchsanweisung-AT-900_01_eng.indd   9

1500_7006000_04260_Gebrauchsanweisung-AT-900_01_eng.indd   9

19.03.2020   07:02:18

19.03.2020   07:02:18

Summary of Contents for AT 870

Page 1: ...niz Switzerland HS Doc no 1500 7006000 04260 26 Edition 2020 03 INSTRUCTIONS FOR USE Applanation tonometer AT 900 AT 870 26 Edition 2020 03 1500_7006000_04260_Gebrauchsanweisung AT 900_01_eng indd 1 1500_7006000_04260_Gebrauchsanweisung AT 900_01_eng indd 1 19 03 2020 07 01 45 19 03 2020 07 01 45 ...

Page 2: ...specially indicated in Glaucoma disease Contraindication There is no absolute contraindication for execution of the tonometry Appropriate professional assessment and caution are necessary The list of known contraindica tions includes but is not limited to active corneal infections corneal epithelium defects eye burns history of recurring epithelial erosion etc WARNING Applanation tonometry may onl...

Page 3: ...eter 7 3 6 Description of models 8 4 Commissioning 8 5 Operation 8 5 1 Astigmatism 8 5 2 How pressure is measured 8 5 3 Preparing the patient 8 5 4 Patient instructions 9 5 5 Preparation of slit lamp and tonometer 9 5 6 Measuring correctly 9 5 7 Sources of error 10 5 7 1 Wrong distance to patient 10 5 7 2 Position too far to the right left 10 5 7 3 Position too high low 11 5 7 4 Incorrect pressure...

Page 4: ...t company Make a report of any damaged parts This report must be signed by you and by the representative of the transport company Leave the device in the packaging for a few hours before unpacking it condensa tion Check the appliance for damage after it is unpacked Return defective appliances in the appropriate packaging Store packaging material carefully so that it can be used for possible return...

Page 5: ...on of diseases to the patient and user as well as damage to the measuring prism Residue from cleaning agents and disinfectants may irritate and burn the patient s eye As a rule the measuring prisms may be prepared together with each other but not with any other products In order to achieve efficient disinfection of tonometer measuring prisms we recommend using the disinfectant Sekusept Forte S It ...

Page 6: ...for the purposes and in the manner des cribed in the documents distributed with the product The product must be treated as described in the Safety chapter Improper hand ling can damage the product This would void all guarantee claims Continued use of a product damaged by incorrect handling may lead to personal injury In such a case the manufacturer will not accept any liability Haag Streit does no...

Page 7: ...n the pivot s pin and swivel the tonometer to the right until it snaps into position The tonometer will remain in this position even when not in use 5 When using the BI 900 BM 900 and BM 900 V the stop peg must be assembled a and when using the BC 900 or BD 900 it must be removed b b a 3 2 AT 900 model T The Applanation Tonometer is tucked into one of the holes on the horizontal guide plate over t...

Page 8: ...E A smooth operation can only be guaranteed with the use of original Haag Streit measuring prisms and Tonosafe 5 Operation WARNING Reusable measuring prisms must be cleaned and disinfected after ev ery examination 5 1 Astigmatism If the cornea is spherical measurements can be taken on any meridian but it is most convenient to measure on the 0 meridian The choice of meridian is of significance howe...

Page 9: ...atient s eyelids should this be necessary for measure ment The illumination of the applanated surface through the measuring prism is practically reflection free Observation with AT 900 model R in the left eyepiece with AT 900 model BQ in the right eyepiece AT 900 model T 1 For an examination through the tonometer s left or right eyepiece the angle between the illumination instrument and the micros...

Page 10: ...he tear fluid dried during the longer lasting measuring Allow the patient to close his eyes a few times and then re peat the measurement 5 7 1 Wrong distance to patient No semi circular image only centre line is visible 3 3 The measuring prism is not coming into contact with the cor nea If the patient draws his head back slightly irregular pulsa tions will be caused because the measuring prism is ...

Page 11: ...the eye eye still too high Using the control lever move the slit lamp upwards Two incomplete circles the larger one on top 13 12 Measuring prism not centred on the eye eye still too high Using the control lever move the slit lamp upwards Correct setting 14 13 Two semicircles appear exactly in the middle of the eyepiece 5 7 4 Incorrect pressure The outside borders of the fluorescein bands are in co...

Page 12: ...2 Maximum limiting deviation 3 Limiting deviation of 0 to 32 66 mN Maximum 0 49 mN 4 Limiting deviation from 32 67 mN Maximum 1 5 of nomi nal value 4 3 0 49 mN 0 49 mN 32 66 mN 0 mN 1 2 2 7 Maintenance WARNING Installation repairs and modifications may only be performed by trained specialists Do not repair or service the device while it is being used on the pa tient 7 1 Repairs To ensure a long se...

Page 13: ...he zero line on the measuring drum is moved downwards opposite the index by a single marking width The feeler arm must move by itself all the way to the stop piece in the direction of the examiner Check position 0 05 The zero line on the measuring drum is moved upwards opposite the index The feeler arm must move all the way to the stop piece on the patient s side 0 05 0 05 Check the drum position ...

Page 14: ... the measuring drum s calibration mark 6 upwards opposite the index by a interval The feeler arm must move towards the patient 5 9 6 1 7 4 Checking the Tonometer AT 870 NOTE This check must be carried out once a month If the check yields errone ous results please check the following points 1 Is the measuring prism properly positioned 2 Is the control weight correctly set 3 Repeat the check Defecti...

Page 15: ... of one marking With light pressure to the feeler arm in the direction of free movement the feeler arm must start to move independently towards the limit stop in the direction of the examiner Check position 2 05 Move calibration mark 2 on the measuring drum upwards opposite the fixed index by a width of one marking The feeler arm must start to move into direction of the stop piece on the patient s...

Page 16: ...eaning and disinfect ing the measuring prism and the separate instructions for use A 3 Tonosafe disposable prisms Tonosafe disposable prisms are a comfortable and effective solution to reduce the risk of cross infection in patients To nosafe is available in sterile packages with 5 supports and 100 disposable prisms B Legal regulations Haag Streit maintains a quality management system in accordance...

Page 17: ...TUGUÊS SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7006000 04260 26 Edition 2020 03 1500_7006000_04260_Gebrauchsanweisung AT 900_01_eng indd 17 1500_7006000_04260_Gebrauchsanweisung AT 900_01_eng indd 17 19 03 2020 07 02 47 19 03 2020 07 02 47 ...

Page 18: ...TUGUÊS SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7006000 04260 26 Edition 2020 03 1500_7006000_04260_Gebrauchsanweisung AT 900_01_eng indd 18 1500_7006000_04260_Gebrauchsanweisung AT 900_01_eng indd 18 19 03 2020 07 02 47 19 03 2020 07 02 47 ...

Page 19: ...TUGUÊS SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7006000 04260 26 Edition 2020 03 1500_7006000_04260_Gebrauchsanweisung AT 900_01_eng indd 19 1500_7006000_04260_Gebrauchsanweisung AT 900_01_eng indd 19 19 03 2020 07 02 47 19 03 2020 07 02 47 ...

Page 20: ...treit com 20 DEUTSCH ENGLISH FRANÇAIS ITALIANO ESPAÑOL NEDERLANDS PORTUGUÊS SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7006000 04260 26 Edition 2020 03 Should you have any further questions please contact your Haag Streit representative at 1500_7006000_04260_Gebrauchsanweisung AT 900_01_eng indd 20 1500_7006000_04260_Gebrauchsanweisung AT 900_01_eng indd 20 19 03 2020 07 02 49 1...

Reviews: