ii.
Take a reading with the oximeter.
iii.
When the Bluetooth icon of the oximeter flashes, remove the finger.
iv.
Turn on the gateway.
v.
The gateway shows the number of currently paired devices with LEDs for 5
seconds.
vi.
The “Readings” LED turns GREEN.
vii.
Wait for about a few minutes.
viii.
At success, the “Readings” LED blinks GREEN 5 times.
ix.
At failure, the “Readings” LED blinks RED 5 times.
x.
The gateway goes to the step “v.”
xi.
You can pair another device at P2 mode successively.
xii.
After completion of pairing, you should turn off, set the mode selection switch
at “N” and turn on the gateway for normal operation.
3)
Pairing
A.
General process
i.
P2 Mode(Master): The same mode as “P2” mode with the mode selection
switch
1.
Go to configuration mode. Please see the section “2. Gateway
configuration” with “setting” checked.
2.
Click “Start” for searching devices. As the result, a list of Bluetooth
devices available is shown in “Device List.”
3.
Select one device and click “Pairing.”
4.
Information about the process will be shown in “Bluetooth Log.”
ii.
P1 Mode(Slave): The same mode as “P1” mode with the mode selection switch
1.
Go to configuration mode. Please see the section “2. Gateway
configuration” with “setting” checked.
2.
Click “Start” to make gateway be pairing standby mode.
3.
Information about the process will be shown in “Bluetooth Log.”
10.
Measurement and data transmission
1)
How to use
A.
Take a reading, and then the reading is transmitted to the gateway by Bluetooth.
B.
And the reading is automatically transmitted to the server via mobile networks.
C.
At server access failure, it tries 9 more times, in total 10 times at most with 1 hour
period until success.
2)
Clock setup
A.
Clock of the gateway is automatically set with 3G connection.
B.
Clocks of measuring devices can be set during communication with the gateway.
C.
But, regarding the IEM’s blood pressure monitor, clock must be set manually.
3)
LED indications
A.
During Bluetooth communication, the “Readings” LED turns GREEN.
B.
At success, the “Readings” LED blinks GREEN 5 times.
C.
At failure, the “Readings” LED blinks RED 5 times.
D.
During server connections, the “Sending” LED turns GREEN.
E.
At success, the “Sending” LED blinks GREEN 5 times.
F.
At failure, the “Sending” LED blinks RED 5 times.
4)
Notification for power failure