![h/p/cosmos cos30018 Original Instructions For Use Download Page 25](http://html.mh-extra.com/html/h-p-cosmos/cos30018/cos30018_original-instructions-for-use_574518025.webp)
disposal
file: \\quasarf\erp$\article\cos14310pw-v1_1_hpc-en\20160706_cos14310pw-v1.1_hpc-en_instruction_for_use_h-p-cosmos_parawalk.doc
© 2016 h/p/cosmos sports & medical gmbh author: fh [email protected] created 06.07.2016 printed 06.07.2016 page: 25 of 30
9
Technical data
9.1
h/p/cosmos parawalk
®
Ask for further details and optional equipment or visit www.h-p-cosmos.com. Subject to technical alterations without prior notice.
9.2
Economic life time
The economic life time of the h/p/cosmos parawalk® is determined at common usage and application to 10 years. Every
maintenance and repair work needs to be carried out by authorized h/p/cosmos technicians. For expendable parts a shorter
life-time is due.
h/p/cosmos parawalk
®
h/
p/
co
sm
os
p
ar
aw
alk
2
,4
4m
h/
p/
co
sm
os
p
ar
aw
alk
3
m
h/
p/
co
sm
os
p
ar
aw
alk
4
m
h/
p/
co
sm
os
p
ar
aw
alk
5
m
order number
cos30018_L2.44
cos30018
cos30019
cos30020
variants
none, wooden plate optional
va01 with wooden plate, va02 without wooden plate
length
244 cm
300 cm
400 cm
500 cm
width (base frame)
102 cm
width (handrails)
450 … 1155 mm (see below)
height
740 … 1075 mm (see below)
weight (va01)
195 kg
210 kg
240 kg
270 kg
weight (va02)
285 kg
300 kg
330 kg
360 kg
max user weight
400 kg
design
white RAL9010, powder coated
Safety standards
c
MDD, directive 93/42/EEC + 2007/47/EC,
tests based on ISO 20957-1 and DIN EN 957-6 where applicable
Field of application / accuracy
medical: S,I (ISO 20957-1)
Classification according to MDD
non-active medical device, risk class I
Environmental conditions:
Transport and storage
temperature: –30...+50 °C, humidity: 0...95 % - without condensation
barometric pressure: 700...1060 hPa
Environmental conditions:
Operation
temperature: +10...+40 °C (deviation on request), humidity: 30...70% - without condensation
(up to 95% on request), barometric pressure: 700...1060 hPa
Maximum operating altitude: approx. 10,000 feet (3000m), without pressurization
handrails
40 mm steel tube
handrail adjustment
gas spring support – very smooth adjustment in height and width with readable scales