17.4 User Options OP (details)
Option Description
Comment / Display
OP01 Reset (delete) error messages
Note: This option exclusively resets the causing
variable. If, for example, the E.02 service-time-
interval is being reset with this operation, the
distance-interval is not reset and vice versa.
Only administrator option A-OP 47 resets all
three values simultaneously.
The required maintenance work must be performed before deleting
error messages. The h/p/cosmos service department must be
contacted before one of the service displays or error codes is deleted!
Confirmation on display
indicating "donE"
Info: This option only resets the error message. If the error still exists,
you are not able to reset the error message. In this case consult an
authorized service engineer
The following intervals are reset by this operation:
E.01: Oil-interval A-OP35
E.02: Service-time-interval A-OP37
E.02: Service-distance-interval A-OP38
OP02 Total distance covered (km)
and
indicates: Total distance covered in km
indicates km
OP03 Indication of total hours of operation
= stand-by-time including runtime of motor/
running belt (h)
and
show: Operation hours
reports: h
OP04 Indication of total hours of
runtime of motor/running belt (h)
and
report: Operation hours
shows: h
OP05 Indication of firmware version and date
reports “OP05”
reports “typE”
indicates device type, e.g. “1.4”
indicates: „MCU 5“
reports version, e.g. “1.01.1”
reports default type, e.g. 1.3
OP06 Adjustment of actual date and the real time clock
shows: rtc for Real Time Clock
reports flashing: Date / time, year, month, date, hours, minutes,
seconds
OP07 Acoustic heart rate signal
This function is normally used to control the regularity of the heart rate
or to find reasons for transmitting problems
indicates: OFF or ON
OFF: No acoustic heart-rate-signal
ON: Acoustic heart rate signal for every beat
OP08 Stop time / deceleration time after STOP key in
relation to the max. speed
indicates: Stop time in sec.
indicates: sec for seconds
Adjustable from 2.... 30 seconds
OP09 Start speed (manual mode or cardio mode) for
feedback after START key has been pressed.
This value can be reduced to 0.0 km/h for
advanced users.
indicates: start speed in km/h
max flashing
set unit is flashing
Adjustable from 0.0 km/h ... 5.0 km/h