page 65 of 99
cos105000_170-190_65, rev. 1.03
© 1988 - 2019 h/p/cosmos sports & medical gmbh
17 Annex I
17.1 Installation
This device must only be unpacked, transported, installed, commissioned, instructed, maintanined and repaired by h/p/cosmos
service personnel (see maintenance).
If the packaging has been damaged, please contact
immediately.
Make sure no objects, sand, stones, liquids, towels, jewellery, cell phones, containers with liquid etc. can fall into the device or onto
the running surface or underneath the running belt.
It is the customer’s responsibility to ensure the following conditions before the installation:
❚
There must be a separate power circuit for the device (dedicated line and fuse).
❚
There must be a separate wall socket for the device (electrically interlocked with circuit breaker 16A, type C).
❚
The wall socket has to be marked with the serial number of the device to ensure it is connected to this socket only.
❚
Use ground wire plugs with tested ground wires only.
❚
The intended location must provide a suitable potential equalization condition (e.g. PE-bolt).
❚
If in compliance with local / national standards, use type A RCDs (residual current operated protective device).
❚
For type B RCDs (measuring also direct current) an additional leakage current limiter has to be installed in the treadmill.
❚
The intended location must provide the requirements for electrical installation acc. to “technical data”.
❚
The intended location must maintain the local requirements for electrical installation.
❚
The wall socket has to be accessible at all times for maintenance and emergency disconnection situations.
❚
The intended location must meet the environmental conditions (see “technical data”).
❚
The intended location must be capable of bearing the load of the device (see “technical data”).
❚
The intended location must provide the safety area and free area as stated in “position of subject and operator”.
❚
The intended location must provide a ceiling height, high enough for accessories (fall prevention).
❚
The intended location must provide a stable, levelled and non-slippery base/floor to prevent noise, bouncing or malfunction.
The manufacturer does not assume liability for any damage, complaints or missing parts that are not reported immediately upon
delivery on the packing list/delivery note.
Disconnect power plug when device ist not used for long periods.