page 62 of 99
cos105000_170-190_65, rev. 1.03
© 1988 - 2019 h/p/cosmos sports & medical gmbh
13.10 Certificates
Description
Data
MD 2006/42/EC
CE declaration of conformity
MD 2006/42/EC
EC certificate, quality assurance
MD 2006/42/EC
Free sales certificate
EN ISO 13485
Certificate, quality management medical devices
Certificates see accompanying documents and
https://www.hpcosmos.com/en/contact-support/media-downloads/certificatesw
13.11 Interfaces (RS232, D-SUB, 9-pole)
Description
Data
stellar [cos30003]
quasar [cos30003]
COM 1 (Baud rate 9600 bps / 115200 bps)
standard
COM 2 (Baud rate 9600 bps )
standard
COM 3 (Baud rate 115200 bps)
service
COM 4 (Baud rate 115200 bps)
n/a (polar)
13.12 Voltage, Current, Performance
Description
Data
stellar [cos30003]
quasar [cos30003]
Input voltage *
230 V ~ (f: 50 / 60 Hz)
Current input (long time) *
7.0 A
Current input (momentary) *
15.2 A
Energy consumption (long time)
≤ 1600 VA
Energy consumption (momentary)
≤ 3500 VA
Energy efficiency
N/A
Device protection switch (circuit breaker)
16 A
Drive motor capacity
3300 W
Elevation motor capacity
No elevation
500 W
Device leakage current
≤ 1.0 mA
Isolation transformer
N/A
Power supply cord
not detachable, 5 m
* Overload or weak power supply may lead to reduced speed accuracy or tripping the fuse.
Subject to technical alterations without prior notice. E & OE (errors and omissions excepted).
Note: In case of special voltage versions, the data on the name plate apply.
Connect mains only if data on nameplate and local power supply are identical.
spor