Gymna V-ACTOR II Accessory Manual Download Page 5

 

Version 252 of 2018-04-06 

 

 

 

CAUTION 

Before using the V-ACTOR II handpiece for the first time, please make sure you 
have read and understood all information provided in this operating manual. 
 
Familiarity  with  the  information  and  instructions  contained  in  this  manual  is 
essential  for  ensuring  efficient  and  optimal  use  of  the  system,  for  avoiding 
hazards to personnel and equipment and for obtaining good treatment results. 
 
Thorough knowledge of the information included in this manual will also enable 
you to react promptly and effectively in the event of malfunctions or failures 
 

 
 

Summary of Contents for V-ACTOR II

Page 1: ...ACCESSORY MANUAL V ACTOR II Handpiece For ShockMaster devices...

Page 2: ......

Page 3: ...How to change the transmitters 6 3 Operation 7 3 1 General warnings and safety information 7 3 2 V ACTOR II treatment parameter settings 7 3 3 Functional checks 8 3 4 Standard operation 8 3 5 Treatme...

Page 4: ...ual contains warnings safety instructions and specific operating instructions in accordance with liability regulations ATTENTION Complete or partial failure to observe the instructions information or...

Page 5: ...Familiarity with the information and instructions contained in this manual is essential for ensuring efficient and optimal use of the system for avoiding hazards to personnel and equipment and for ob...

Page 6: ...ptional accessory for the ShockMaster control units which may be used to complement the radial shockwave therapy The V ACTOR II handpiece allows the treatment of soft tissue with high frequency pulses...

Page 7: ...pment transport 2 2 Scope of the supply V ACTOR II with 2 applicators 25mm and 40mm 2 3 Handpiece connection NOTE Please also refer to the separate operating instructions of your control unit Connect...

Page 8: ...on is initially released then the plug is pulled out of the handpiece connector 2 2 3 1 Compatibility The V ACTOR II handpiece can be connected to the ShockMaster control units from the manufacturer G...

Page 9: ...onnect to the ShockMaster control units 3 2 V ACTOR II treatment parameter settings After connecting the V ACTOR II handpiece the control unit automatically recognizes V ACTOR II and operating mode wi...

Page 10: ...he V ACTOR II handpiece to the control unit Set the pressure to an initial value of 2 0 bar and a frequency of 20Hz Start the V ACTOR II treatment by pushing the trigger button 3 5 Treatment CAUTION A...

Page 11: ...r any other agent suitable for surface disinfection that does not harm the material lacquered sheet metal In case of doubt please consult an accepted list of disinfectants e g the VAH list of disinfec...

Page 12: ...eed in accordance with national regulations At the end of its life time the V ACTOR II must be disposed of as electronic waste 4 2 Repair Repair work on defective equipment must only be carried out by...

Page 13: ...ions No compressed air supply Leaks on handpiece cable or cable not properly connected Check the cable connection and replace them if necessary Check compressor see User Manual control device No pulse...

Page 14: ...nd spare parts 6 1 V ACTOR II handpiece ShockMaster V ACTOR II BLACK 338 239 ShockMaster V ACTOR 25mm Applic Black 338 668 ShockMaster V ACTOR 40mm Applic Black 338 679 NOTE For information on various...

Page 15: ...0 0 5 35 Hz Compressed air input ShockMaster 500 1 5 bar ShockMaster 300 1 4 bar Operating conditions Ambient temperature 10 C 35 C Atmospheric pressure 70 106 kPa Relative humidity 30 75 Storage and...

Page 16: ...ned products shall be borne by the customer Warranty claims will only be accepted if the handpiece is returned in its complete and original state cleaned and in the case without accessories Missing co...

Page 17: ...Version 252 of 10 04 2018...