5. General technical data.
•
Treatment unit
Mains voltage:
115/230 Vac +- 10%
Frequency:
50/60
Hz
Mains fuses:
1A type L
Current consumption:
maximum 500 mA
Dimensions
413
✕
280
✕
100 mm
Weight:
5
kg.
Labels
Hazard warning label
Serial number label
Type identification label
Fuse value label
Bauartzulassungs
Nr.
Manufacturer, mains, Voltage
Frequency and power consumption
Class according to IEC601-1:
I
Type according to IEC601-1:
BF
Limits test values according to IEC 601-1:
Earth leakage current: IEC requirement < 500
µ
A
Earth leakage single fault: IEC requirement < 1 mA
Enclosure leakage current normal: IEC requirement < 100
µ
A
Enclosure leakage single fault: IEC requirement < 500
µ
A
Patient leakage normal: IEC requirement < 100 UA
Patient leakage single fault: IEC requirement < 500
µ
A
Patient leakage mains on applied part sfc: IEC requirement < 5
mA
Patient auxiliary current normal: IEC requirement < 10
µ
A
Patient auxiliary single fault: IEC requirement < 50
µ
A
Earth resistance: typical 0.12
Ω
Number of channels:
2
Continuous use:
Yes
Useable in explosive environment:
No
•
Treatment heads
1 Mhz, large:
Surface: 5 cm²
Max output power: 10 W
3 Mhz, small:
Surface: 0.8 cm²
Max output power: 1.6 W
Page 6