OPERATION
31
2.
Set the tracking switch
position to Series, + .
TRAC
KING
INDE
P.
SERI
ES
PARA
LLEL
3.
Set the OVP if necessary. In
tracking series mode, set the
Channel2 (Slave) OVP setting to
the maximum level, so that the
OVP trips if the Channel1
(Master) setting is violated. For
OVP setup details, see page22.
O.V.P.
SET
TRIP
4.
Connect the load to the front panel terminals,
c & channel2
−
(Single supply).
CH2
CH1
GND
0 60
V , 3A
0 30
V , 6A
0 60
V , 3A
0 30
V , 6A
5.
Turn up the Channel2
Current knob to maximum.
CURRENT
CH2
SLAVE
6.
Set the output Voltage and
Current using the Channel1
(Master) knob.
VOLTAGE
CURRENT
CH1
MASTER
C.V.
C.C.
SPD-3606 User Manual
32
7.
Refer to the Channel1 (Master) meter and
indicator for the output setting level and
CV/CC status.
CH1 / CH3
CH1
MASTER
C.V.
C.C.
Current
level
Channel1 meter reading shows
the output Current. (Channel2
Current control must be in the
Maximum position).
Voltage
level
Double the reading on the
Channel1 Voltage meter. (In the
above case, the actual output is
23.6 x 2 = 47.2V).
8.
Press the Output key. The
Output indicator turns
green.
OUTPUT
ON / OFF
OUTPUT
ON / OFF